Your browser doesn't support javascript.
loading
Biomechanical evaluation comparing zero-profile devices versus fixed profile systems in a cervical hybrid decompression model: a biomechanical in vitro study.
Kinon, Merritt D; Greeley, Samantha L; Harris, Jonathan A; Gelfand, Yaroslav; Yassari, Reza; Nakhla, Jonathan; De la Garza-Ramos, Rafael; Patel, Pavan; Mirabile, Belin; Bucklen, Brandon S.
Affiliation
  • Kinon MD; Department of Neurological Surgery, Montefiore Medical Center, Bronx, NY, USA.
  • Greeley SL; Musculoskeletal Education and Research Center, Globus Medical, Inc., Audubon, PA, USA. Electronic address: sgreeley@globusmedical.com.
  • Harris JA; Musculoskeletal Education and Research Center, Globus Medical, Inc., Audubon, PA, USA.
  • Gelfand Y; Department of Neurological Surgery, Montefiore Medical Center, Bronx, NY, USA.
  • Yassari R; Department of Neurological Surgery, Montefiore Medical Center, Bronx, NY, USA.
  • Nakhla J; Department of Neurological Surgery, Montefiore Medical Center, Bronx, NY, USA.
  • De la Garza-Ramos R; Department of Neurological Surgery, Montefiore Medical Center, Bronx, NY, USA.
  • Patel P; School of Biomedical Engineering, Science and Health Systems, Drexel University, Philadelphia, PA, USA.
  • Mirabile B; Department of Mechanical Engineering, College of Engineering, University of Notre Dame, Notre Dame, IN, USA.
  • Bucklen BS; Musculoskeletal Education and Research Center, Globus Medical, Inc., Audubon, PA, USA.
Spine J ; 20(4): 657-664, 2020 04.
Article in En | MEDLINE | ID: mdl-31634616
ABSTRACT
BACKGROUND CONTEXT The use of zero-profile devices and the need for posterior fixation in conjunction with a cervical hybrid decompression model have yet to be investigated.

PURPOSE:

To compare the biomechanics of zero-profile and fixed profile cervical hybrid constructs composed of anterior cervical discectomy and fusion (ACDF) and anterior cervical corpectomy and fusion (ACCF). Fixed profile devices included anterior plating, whereas zero-profile devices included integrated screws. STUDY

DESIGN:

In vitro cadaveric biomechanical study.

METHODS:

Twelve fresh-frozen cadaveric spines (C2-C7) were divided into two groups of equal bone mineral density, fixed profile versus zero profile (n=6). Groups were instrumented from C3-C6 with either (1) an expandable ACCF device and a static ACDF spacer with an anterior plate (Hybrid-AP) or (2) a zero-profile ACCF spacer with adjacent zero-profile ACDF spacer (Hybrid-Z). Motion was captured for the (1) intact condition, (2) a hybrid model with lateral mass screws (LMS), (3) a hybrid model without LMS, and (4) a hybrid model without LMS following simulated repetitive loading (fatigue).

RESULTS:

Hybrid-AP with LMS reduced motion in flexion-extension (FE), lateral bending (LB), and axial rotation (AR) by 77%, 88%, and 82%, respectively, compared with intact. Likewise, Hybrid-Z with LMS exhibited the greatest reduction in motion relative to intact in FE, LB, and AR by 90%, 95%, and 66%, respectively. Following simulated in vivo fatiguing, an increase in motion was observed for both groups in all planes, particularly during Hybrid-Z postfatigue condition where motion increased relative to intact by 29%. Overall, biomechanical equivalency was observed between Hybrid-AP and Hybrid-Z groups (p>.05). Three (50%) of the Hybrid-Z group specimens exhibited signs of implant migration from the inferior endplate during testing.

CONCLUSIONS:

Fixed profile systems using an anterior plate for supplemental fixation is biomechanically more favorable to maintain stability and prevent dislodgement. Dislodgement of 50% of the Hybrid-Z group without LMS emphasizes the necessity for posterior fixation in a zero-profile cervical hybrid decompression model.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Spinal Fusion / Cervical Vertebrae Limits: Humans Language: En Journal: Spine J Journal subject: ORTOPEDIA Year: 2020 Document type: Article Affiliation country: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Spinal Fusion / Cervical Vertebrae Limits: Humans Language: En Journal: Spine J Journal subject: ORTOPEDIA Year: 2020 Document type: Article Affiliation country: United States