Antimicrobial Susceptibility Test for the Determination of Resistant and Susceptible S. aureus and Enterococcus spp. Using a Multi-Channel Surface Plasmon Resonance Device.
Diagnostics (Basel)
; 9(4)2019 Nov 15.
Article
in En
| MEDLINE
| ID: mdl-31731591
The objective of this study was to investigate the development of a surface plasmon resonance (SPR) sensor platform equipped with multiple channels for the simultaneous determination of methicillin-resistant S. aureus (MRSA), methicillin-susceptible S. aureus (MSSA) and vancomycin-resistant Enterococcus (VRE), and vancomycin-susceptible Enterococcus (VSE). Drug resistance of S. aureus strains against cefoxitin and Enterococcus strains against vancomycin were investigated both using the minimum inhibitory concentration method (MIC) assay and the SPR system equipped with single and multiple channels. The MIC values of MRSA and MSSA ranged from 32 µg/mL to >128 µg/mL and from 1 µg/mL to 4 µg/mL, respectively. The MIC values of VRE and VSE were between 64 to >128 µg/mL and 2-4 µg/mL, respectively. With the multiple-channel system, the angle shifts of MRSA, MSSA, VRE and VSE were found to be -0.030° and -0.260°, -0.010° and -0.090° respectively. The antibiotic-resistant and susceptible strains were distinguished within 3 h for S. aureus strains and within 6 h for Enterococcus strains.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Language:
En
Journal:
Diagnostics (Basel)
Year:
2019
Document type:
Article
Affiliation country:
Turkey
Country of publication:
Switzerland