Comparison of Leadless Pacing and Temporary Externalized Pacing Following Cardiac Implanted Device Extraction.
Gonzales, Holly; Richardson, Travis D; Montgomery, Jay A; Crossley, George H; Ellis, Christopher R.
Affiliation
  • Gonzales H; Cardiovascular Division, Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, TN, USA.
  • Richardson TD; Cardiovascular Division, Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, TN, USA.
  • Montgomery JA; Cardiovascular Division, Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, TN, USA.
  • Crossley GH; Cardiovascular Division, Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, TN, USA.
  • Ellis CR; Cardiovascular Division, Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, TN, USA.
J Innov Card Rhythm Manag ; 10(12): 3930-3936, 2019 Dec.
Article in En | MEDLINE | ID: mdl-32477715
Pacemaker-dependent (PD) patients undergoing implantable cardiac electronic device extraction often must be subjected to temporary pacing interventions. We sought to determine the safety and utility of a leadless pacing system (Micra™; Medtronic, Minneapolis, MN, USA) in patients undergoing system extraction as compared with externalized temporary transvenous right ventricular lead (temp-perm) placement. We performed a retrospective cohort analysis of all patients receiving either permanent Micra™ or temp-perm systems following system extraction from October 2013 to September 2017 at Vanderbilt University Hospital. The Micra™ and temp-perm cohorts included nine and 27 patients meeting the inclusion criteria, respectively. System infection was the most common indication for extraction (67% Micra™, 84% temp-perm), but no patients had active bacteremia at the time of permanent system reimplantation. There was no difference in system type (p = 0.09) or mean lead dwell time extracted (109 versus 81 months; p = 0.93). Procedure times were comparable between the two groups (180 versus 194 minutes; p = 0.74). Patients receiving Micra™ systems had shorter hospital stays after extraction (two versus eight days; p < 0.005), with no difference in major complications (11% versus 15%; p = 0.78) or 30-day (11% versus 7%; p = 0.77) or 90-day (11% versus 11%; p = 0.45) mortality. No reinfections were observed in either group at 90 days. Implantation of the Micra™ pacing system in select PD patients after system extraction is feasible and appears to reduce the hospital length of stay as compared with the use of temp-perm systems.