Your browser doesn't support javascript.
loading
Performance characteristics and optimal cut-off value of triple adenylate nucleotides test versus adenosine triphosphate test as point-of-care testing for predicting inadequacy of duodenoscope reprocessing.
Ridtitid, W; Pakvisal, P; Chatsuwan, T; Kerr, S J; Piyachaturawat, P; Luangsukrerk, T; Kongkam, P; Rerknimitr, R.
Affiliation
  • Ridtitid W; Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand; Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Tha
  • Pakvisal P; Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Chatsuwan T; Department of Microbiology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand; Antimicrobial Resistance and Stewardship Research Unit, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
  • Kerr SJ; Biostatistics Excellence Centre, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
  • Piyachaturawat P; Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Luangsukrerk T; Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Kongkam P; Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand; Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Tha
  • Rerknimitr R; Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand; Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Tha
J Hosp Infect ; 106(2): 348-356, 2020 Oct.
Article in En | MEDLINE | ID: mdl-32768608
ABSTRACT

BACKGROUND:

Adenosine triphosphate (ATP) test based on one nucleotide has been applied as point-of-care testing (POCT) for bacterial contamination in the medical and food industries. Hypothetically, testing three adenylate nucleotides (A3) may provide better detection of duodenoscope bacterial contamination than ATP test.

AIM:

To evaluate performance characteristics and optimal cut-off value of A3 and ATP tests in predicting bacterial contamination of duodenoscopes.

METHODS:

Four hundred duodenoscope samples obtained after 100 endoscopic retrograde cholangiopancreatography procedures were randomized into group A (A3 test) or B (ATP test). Samples were collected from the elevator at the four-step cleaning process of duodenoscope. We defined the new cut-off value of the test for reaching 100% negative predictive value (NPV) from our receiver operating characteristic (ROC).

FINDINGS:

Using the cultures from the four-step cleaning process as the reference, the areas under ROC (AUROC) were 0.83 and 0.84 for group A (N = 200) and group B (N = 200), respectively. Using the cultures from post-high-level disinfection (HLD) as the reference, the AUROC were 0.35 and 0.74 for group A (N = 50) and group B (N = 50), respectively. We investigated ATP as a POCT after HLD with a new cut-off value of 40 RLU. However, this threshold did not allow detection of low numbers of bacteria.

CONCLUSION:

A3 and ATP tests provide good performances in predicting bacterial contamination of duodenoscopes for the four-step cleaning process. The ATP <40 RLU is helpful as a POCT after HLD; however, the limitation of this cut-off value is its inability to detect low numbers of bacteria.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Bacteria / Adenosine Triphosphate / Disinfection / Duodenoscopes / Point-of-Care Testing / Nucleotides Type of study: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies / Risk_factors_studies Limits: Humans Language: En Journal: J Hosp Infect Year: 2020 Document type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Bacteria / Adenosine Triphosphate / Disinfection / Duodenoscopes / Point-of-Care Testing / Nucleotides Type of study: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies / Risk_factors_studies Limits: Humans Language: En Journal: J Hosp Infect Year: 2020 Document type: Article