Your browser doesn't support javascript.
loading
A method for spectral irradiance measurement based on a large area WC-C fixed point blackbody.
Opt Express ; 28(19): 28430-28440, 2020 Sep 14.
Article in En | MEDLINE | ID: mdl-32988113
ABSTRACT
We have proposed a novel method for the spectral irradiance measurement of the standard lamp. The measurement has been realized by using national primary scale based on a large area WC-C fixed point blackbody for the first time. The wavelength range is from 450 nm to 1000 nm. The spectral irradiance of the standard lamp has been measured based on the traditional variable temperature blackbody and large area WC-C fixed point blackbody. The results of the two methods agree within the uncertainties. Compared with the traditional measurement method, the fixed point method can reduce the major component of the spectral irradiance measurement uncertainty related to the temperature measurement.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Opt Express Journal subject: OFTALMOLOGIA Year: 2020 Document type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Opt Express Journal subject: OFTALMOLOGIA Year: 2020 Document type: Article