Your browser doesn't support javascript.
loading
How do the geometries of the broach handles relate to the distribution of force and moments in a femoral model?
Fonseca Ulloa, Carlos A; Seeger, Anja; Ishaque, Bernd A; Ahmed, Gafar A; Rickert, Markus; Jahnke, Alexander.
Affiliation
  • Fonseca Ulloa CA; Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany. Electronic address: Carlos.A.Fonseca-Ulloa@med.uni-giessen.de.
  • Seeger A; Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany.
  • Ishaque BA; Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany; Department of Orthopaedics and Orthopaedic Surgery, University Hospital Giessen and Marburg (UKGM), Klinikstrasse 33, 35392 Giessen, Germany.
  • Ahmed GA; Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany; Department of Orthopaedics and Orthopaedic Surgery, University Hospital Giessen and Marburg (UKGM), Klinikstrasse 33, 35392 Giessen, Germany.
  • Rickert M; Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany; Department of Orthopaedics and Orthopaedic Surgery, University Hospital Giessen and Marburg (UKGM), Klinikstrasse 33, 35392 Giessen, Germany.
  • Jahnke A; Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany.
Med Eng Phys ; 86: 122-127, 2020 12.
Article in En | MEDLINE | ID: mdl-33261725
The continuous improvement of minimally invasive hip endoprostheses surgery comes with a change in geometries of surgery instruments like the broach handles. Consequently, depending on the handles' curvature this results in a deviation between handle and femoral axis. Therefore, this study aimed to prove the influence of different handles' curvatures on the preparation of implant site and acting forces and moments in this process. Five femoral models attached to different handles (double-curved, single-curved, straight) were locked in a drop-weight device with standardize implantation forces and moments and five strokes were measured for each possible combination. Distribution of force and moment components was dependent on the handle's curvature, where the lowest variation from the standard force values was by the straight one (av:15.2% ± 0.5%) and the strongest discrepancies were exhibit by the double-curved one (av:54.3% ± 0.1%.). Moment values have also shown this trend with the lowest variation (12.4%-23.3%) by the straight one and the highest discrepancies (56,6%-90.9%) by the double-curved one. Results show that unguided axial impact introduces unwanted transverse forces and moments into the femur. Therefore, broach handles should be modified accordingly so that minimally invasive surgery remains feasible but unwanted forces or moments can still be compensated.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Mechanical Phenomena / Femur Limits: Humans Language: En Journal: Med Eng Phys Journal subject: BIOFISICA / ENGENHARIA BIOMEDICA Year: 2020 Document type: Article Country of publication: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Mechanical Phenomena / Femur Limits: Humans Language: En Journal: Med Eng Phys Journal subject: BIOFISICA / ENGENHARIA BIOMEDICA Year: 2020 Document type: Article Country of publication: United kingdom