Your browser doesn't support javascript.
loading
A porcine study of the area of heated tissue during hot-balloon ablation: Implications for the clinical efficacy and safety.
Nakahara, Shiro; Wakamatsu, Yuji; Sato, Hirotsugu; Otsuka, Naoto; Fukuda, Reiko; Watanabe, Ryuta; Kurokawa, Sayaka; Ishikawa, Tetsuya; Takaoka, Motoki; Nagashima, Koichi; Kobayashi, Sayuki; Taguchi, Isao; Okumura, Yasuo.
Affiliation
  • Nakahara S; Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Koshigaya, Japan.
  • Wakamatsu Y; Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Sato H; Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Koshigaya, Japan.
  • Otsuka N; Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Fukuda R; Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Koshigaya, Japan.
  • Watanabe R; Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Kurokawa S; Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Ishikawa T; Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Koshigaya, Japan.
  • Takaoka M; Hot-Balloon Catheter Business Department, Product Development Group, Toray Industries, Inc, Tokyo, Japan.
  • Nagashima K; Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Kobayashi S; Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Koshigaya, Japan.
  • Taguchi I; Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Koshigaya, Japan.
  • Okumura Y; Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
J Cardiovasc Electrophysiol ; 32(2): 260-269, 2021 02.
Article in En | MEDLINE | ID: mdl-33382509
ABSTRACT

INTRODUCTION:

Hot-balloon ablation depends solely on thermal conduction, and myocardial tissue is ablated by only conductive heating from the balloon surface. Despite growing clinical evidence of the efficacy and safety of hot-balloon ablation for atrial fibrillation (AF), the actual tissue temperature and the mechanism of heating during such ablation has not been clarified. To determine, by means of a porcine study, the temperatures of tissues targeted during hot-balloon ablation of AF performed with hot-balloon set temperatures of 73°C or 70°C, in accordance with the temperatures now used clinically.

METHODS:

After a right thoracotomy, thermocouples with markers were implanted epicardially on the superior vena cava (SVC) and pulmonary veins (PVs) in six pigs. The tissue temperatures during hot-balloon ablation (balloon set temperatures of 73°C and 70°C, 180 s/PV) were recorded, and the maximum tissue temperatures and fluoroscopically measured distance from the balloon surface to the target tissues were assessed.

RESULTS:

Sixteen SVC- and 18 PV-targeted energy deliveries were performed. Full-thickness circumferential PV lesions were created with all hot-balloon applications. A significant inverse relation was found between the recorded tissue temperatures and distance (r = -.67; p < .001) from the balloon surface. No tissue temperature exceeded either of the balloon set temperatures. The best distance cutoff value for achieving lethal tissue temperatures more than 50°C was 3.6 mm.

CONCLUSION:

The hot-balloon set temperature, energy delivery time, and tissue temperature data obtained in this porcine study supported the clinical efficacy and safety of the hot-balloon ablation as currently practiced in patients with AF.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pulmonary Veins / Atrial Fibrillation / Catheter Ablation Type of study: Diagnostic_studies Limits: Animals / Humans Language: En Journal: J Cardiovasc Electrophysiol Journal subject: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Year: 2021 Document type: Article Affiliation country: Japan

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pulmonary Veins / Atrial Fibrillation / Catheter Ablation Type of study: Diagnostic_studies Limits: Animals / Humans Language: En Journal: J Cardiovasc Electrophysiol Journal subject: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Year: 2021 Document type: Article Affiliation country: Japan