Your browser doesn't support javascript.
loading
Technical Note: Practical implementation strategies of cycloidal computed tomography.
Roche I Morgó, Oriol; Vittoria, Fabio; Endrizzi, Marco; Olivo, Alessandro; Hagen, Charlotte K.
Affiliation
  • Roche I Morgó O; Department of Medical Physics and Biomedical Engineering, University College London, London, WC1E 6BT, UK.
  • Vittoria F; Department of Medical Physics and Biomedical Engineering, University College London, London, WC1E 6BT, UK.
  • Endrizzi M; ENEA Radiation Protection Institute, Bologna, Italy.
  • Olivo A; Department of Medical Physics and Biomedical Engineering, University College London, London, WC1E 6BT, UK.
  • Hagen CK; Department of Medical Physics and Biomedical Engineering, University College London, London, WC1E 6BT, UK.
Med Phys ; 48(10): 6524-6530, 2021 Oct.
Article in En | MEDLINE | ID: mdl-34169514
ABSTRACT

PURPOSE:

Cycloidal computed tomography is a novel imaging concept which combines a highly structured x-ray beam, offset lateral under-sampling, and mathematical data recovery to obtain high-resolution images efficiently and flexibly, even with relatively large source focal spots and detector pixels. The method reduces scanning time and, potentially, delivered dose compared to other sampling schemes. This study aims to present and discuss several implementation strategies for cycloidal computed tomography (CT) in order to increase its ease of use and facilitate uptake within the imaging community.

METHODS:

The different implementation strategies presented are step-and-shoot, continuous unidirectional, continuous back-and-forth, and continuous pixel-wise scanning. In step-and-shoot scans the sample remains stationary while frames are acquired, whereas in all other cases the sample moves through the scanner continuously. The difference between the continuous approaches is the trajectory by which the sample moves within the field of view.

RESULTS:

All four implementation strategies are compatible with a standard table-top x-ray setup. With the experimental setup applied here, step-and-shoot acquisitions yield the best spatial resolution (around 30 µm), but are the most time-consuming (1.4 h). Continuous unidirectional and back-and-forth images have resolution between 30 and 40 µm, and are faster (35 min). Continuous pixel-wise images are equally time-efficient, although technical challenges caused a small loss in image quality with a resolution of about 50 µm.

CONCLUSION:

The authors show that cycloidal CT can be implemented in a variety of ways with high quality results. They believe this posits cycloidal CT as a powerful imaging alternative to more time-consuming and less flexible methods in the field.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tomography, X-Ray Computed Type of study: Diagnostic_studies Language: En Journal: Med Phys Year: 2021 Document type: Article Affiliation country: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tomography, X-Ray Computed Type of study: Diagnostic_studies Language: En Journal: Med Phys Year: 2021 Document type: Article Affiliation country: United kingdom