A Miniature, Fiber-Optic Vibrometer for Measuring Unintended Acoustic Output of Active Hearing Implants during Magnetic Resonance Imaging.
Sensors (Basel)
; 21(19)2021 Oct 02.
Article
in En
| MEDLINE
| ID: mdl-34640909
Making use of magnetic resonance imaging (MRI) for diagnostics on patients with implanted medical devices requires caution due to mutual interactions between the device and the electromagnetic fields used by the scanner that can cause a number of adverse events. The presented study offers a novel test method to quantify the risk of unintended output of acoustically stimulating hearing implants. The design and operating principle of an all-optical, MRI safe vibrometer is outlined, followed by an experimental verification of a prototype. Results obtained in an MRI environment indicate that the system can detect peak displacements down to 8 pm for audible frequencies. Feasibility testing was performed with an active middle ear implant that was exposed to several pulse sequences in a 1.5 Tesla MRI environment. Magnetic field induced actuator vibrations, measured during scanning, turned out to be equivalent to estimated sound pressure levels between 25 and 85 dB SPL, depending on the signal frequency. These sound pressure levels are situated well below ambient sound pressure levels generated by the MRI scanning process. The presented case study therefore indicates a limited risk of audible unintended output for the examined hearing implant during MRI.
Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Acoustics
/
Magnetic Resonance Imaging
Limits:
Humans
Language:
En
Journal:
Sensors (Basel)
Year:
2021
Document type:
Article
Affiliation country:
Belgium
Country of publication:
Switzerland