Membrane-based lateral flow immunochromatographic strips (ICSs) are useful tools for low-cost self-diagnosis and have been efficiently applied to toxin, physiological index and clinical biomarker detection. In this protocol, we provide a detailed description of the steps to develop a rapid, sensitive and quantitative lateral-flow immunoassay (using AuNPs as a marker and mAbs as a probe). The procedure describes the preparation and characterization of colloidal gold, synthesis of the AuNP-mAb conjugate, assembly of the immunochromatographic strip, and methodological investigation of the assay. The results showed that the final strips can be further utilized for the rapid and convenient self-diagnosis of a small molecule, which may provide an alternative tool in the rapid and precise analysis of physiological and biological indices.