Your browser doesn't support javascript.
loading
Clinical assessment of image quality, usability and patient comfort in dedicated spiral breast computed tomography.
Schmidt, Cynthia Sabrina; Zellweger, Caroline; Wieler, Jann; Berger, Nicole; Marcon, Magda; Frauenfelder, Thomas; Boss, Andreas.
Affiliation
  • Schmidt CS; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland. Electronic address: cynthiasabrina.schmidt@uzh.ch.
  • Zellweger C; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland.
  • Wieler J; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland.
  • Berger N; Institute of Radiology, Spital Lachen, Oberdorfstrasse 41, CH-8853 Lachen, Switzerland.
  • Marcon M; Department of Radiology, Ente Ospedaliero Cantonale (EOC), Via Tesserete 46, CH-6900 Lugano, Switzerland.
  • Frauenfelder T; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland.
  • Boss A; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland.
Clin Imaging ; 90: 50-58, 2022 Oct.
Article in En | MEDLINE | ID: mdl-35917662
ABSTRACT

OBJECTIVE:

To investigate aspects of image quality, feasibility and patient comfort in dedicated spiral breast computed tomography (B-CT) in a large patient cohort.

METHODS:

This retrospective study was approved by the institutional review board. 2418 B-CT scans from 1222 women examined between 04/16/2019 and 04/13/2022 were analyzed. Patients evaluated their comfort during the examination, radiographers carrying out the scans evaluated the patient's mobility and usability of the B-CT device, whereas radiologists assessed lesion contrast, detectability of calcifications, breast coverage and overall image quality. For semi-quantitative assessment, a Likert-Scale was used and statistical significance and correlations were calculated using ANOVAs and Spearman tests.

RESULTS:

Comfort, mobility and usability of the B-CT were rated each with either "no" or "negligible" complaints in >99%. Image quality was rated with "no" or "negligible complaints" in 96.7%. Lesion contrast and detectability of calcifications were rated either "optimal" or "good" in 92.6% and 98.4%. "Complete" and "almost complete" breast coverage were reported in 41.9%, while the pectoral muscle was found not to be covered in 56.0%. Major parts of the breast were not covered in 2.1%. Some variables were significantly correlated, such as age with comfort (ρ = -0.168, p < .001) and mobility (ρ = -0.172, p < .001) as well as patient weight with lesion contrast (ρ = 0.172, p < .001) and breast coverage (ρ = -0.109, p < .001).

CONCLUSIONS:

B-CT provides high image quality and contrast of soft tissue lesions as well as calcifications, while covering the pre-pectoral areas of the breast remains challenging. B-CT is easy to operate for the radiographer and comfortable for the majority of women.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Calcinosis / Mammography Type of study: Observational_studies Limits: Female / Humans Language: En Journal: Clin Imaging Journal subject: DIAGNOSTICO POR IMAGEM Year: 2022 Document type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Calcinosis / Mammography Type of study: Observational_studies Limits: Female / Humans Language: En Journal: Clin Imaging Journal subject: DIAGNOSTICO POR IMAGEM Year: 2022 Document type: Article