Evaluating the effect of Novarerus NV800 air purifier units during orthopaedic surgery to reduce bioburden in the air.
Lytsy, B; Ljungqvist, B; Nordenadler, J; Reinmüller, B.
Affiliation
  • Lytsy B; Clinical Microbiology and Infection Control, Department of Medical Sciences, Uppsala University, Uppsala, Sweden. Electronic address: birgitta.lytsy@akademiska.se.
  • Ljungqvist B; Building Services Engineering, Chalmers University of Technology, Göteborg, Sweden.
  • Nordenadler J; Clinical Science, Intervention and Technology, Function and Technology, Karolinska Institute, Solna, Sweden.
  • Reinmüller B; Building Services Engineering, Chalmers University of Technology, Göteborg, Sweden.
J Hosp Infect ; 130: 108-111, 2022 Dec.
Article in En | MEDLINE | ID: mdl-36113635
BACKGROUND: A locally installed air purifier unit (Novaerus Protect 800) has been shown to reduce the air bioburden in an intensive care unit and the incidence of healthcare-associated infections. AIM: To explore whether this type of air purifying unit could reduce bacterial concentrations in the air of an operating room (OR) during orthopaedic surgery, thereby reducing the risk of surgical site infections. METHODS: In this prospective experimental study, undertaken in 2018, three air purifying units were installed in an OR in a Swedish hospital in 2018. The air was actively sampled during 11 operations by a slit-to-slit agar impactor with the air purifying units either switched on or switched off. Air movements were visualized with the aid of smoke in mock-up studies. FINDINGS: No significant difference in bacterial concentrations in air was found between the two conditions (air purifying units switched off or on) (P=0.54). Air movements around and above the surgical wound were disordered and resembled those of dilution mixing air. CONCLUSION: The three air purifying units installed in the OR did not reduce the airborne bacterial levels in the critical zone during orthopaedic surgery.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Orthopedic Procedures / Air Filters Type of study: Observational_studies / Risk_factors_studies Limits: Humans Language: En Journal: J Hosp Infect Year: 2022 Document type: Article Country of publication: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Orthopedic Procedures / Air Filters Type of study: Observational_studies / Risk_factors_studies Limits: Humans Language: En Journal: J Hosp Infect Year: 2022 Document type: Article Country of publication: United kingdom