A Novel Approach to Neuroendovascular Treatment Looking Up from the Ascending Aorta Using an Alpha Shape Guiding Catheter.
Ohshima, Tomotaka; Ioku, Tetsuya; Yokota, Mao; Kawaguchi, Reo; Matsuo, Naoki; Miyachi, Shigeru.
Affiliation
  • Ohshima T; Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi, Japan. Electronic address: tmtkoh@gmail.com.
  • Ioku T; Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi, Japan.
  • Yokota M; Department of Neurosurgery, Aichi Medical University, Nagakute, Aichi, Japan.
  • Kawaguchi R; Department of Neurosurgery, Aichi Medical University, Nagakute, Aichi, Japan.
  • Matsuo N; Department of Neurosurgery, Aichi Medical University, Nagakute, Aichi, Japan.
  • Miyachi S; Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi, Japan; Department of Neurosurgery, Aichi Medical University, Nagakute, Aichi, Japan.
World Neurosurg ; 169: 32-35, 2023 01.
Article in En | MEDLINE | ID: mdl-36328168
ABSTRACT

BACKGROUND:

A method of guiding an intermediate catheter from a new alpha-type guiding catheter placed in the ascending aorta to the carotid artery is evaluated in neuroendovascular treatment in challenging anatomic directions of the guiding catheter, such as the bovine aortic arch and type III aortic arch.

METHODS:

The existing 8-Fr guiding catheter was given a strong bending shape at the tip to make it an alpha type. The total length of the catheter was 85 cm. This guiding catheter was inserted into the ascending aorta to verify whether a 6-Fr intermediate or aspiration catheter could be coaxially guided into the right and left internal carotid arteries. A silicone vascular model was used for evaluation and in actual clinical cases.

RESULTS:

Creating an alpha shape of the catheter at the aortic arch was very easy. The inner catheter could be easily guided from the brachiocephalic artery to the right common carotid artery by pushing the alpha shape guiding catheter toward the aortic valve. The catheter was easily guided into the left common carotid artery when the α-guide was pulled a little bit backward. The 0.071-inch lumen aspiration catheter reached the bilateral middle cerebral arteries.

CONCLUSIONS:

The 8-Fr alpha shape guiding catheter quickly guides the inner catheter into the bovine and the type III aortic arch by looking up from the ascending aorta.