Accuracy of pilot balloon palpation for cuff pressure assessment in small versus large sized tubes: a prospective non-randomized observational study.
Sci Rep
; 13(1): 5580, 2023 04 05.
Article
in En
| MEDLINE
| ID: mdl-37020098
Pilot balloon palpation is still a commonly used method to evaluate cuff pressure of the endotracheal tube after intubation. This study determined whether the size of the tracheal tube influenced the accuracy of pilot balloon palpation. A prospective observational analysis of 208 patients intubated with an endotracheal tube of internal diameter (ID) 6.0 or 8.0 was conducted. An anesthesiologist judged the cuff pressure by manual pilot balloon palpation, and then measured the cuff pressure with a pressure gauge. Cuff pressure exceeding 20-30 cmH2O was defined as false recognition. The intracuff pressure was significantly higher in ID 6.0 tube than in the ID 8.0 tube (41.9 ± 18.8 cmH2O vs. 30.3 ± 11.9 cmH2O, p < 0.001). The number of patients that were mistakenly perceived to have appropriate cuff pressure by pilot balloon palpation was significantly higher in the ID 6.0 group compared to the ID 8.0 group (85 (81.7%) vs. 64 (61.5%), p = 0.001). Therefore, a smaller tube size may further increase risk of inaccurate measurement by pilot balloon palpation and although pressure gauge is recommended for all sizes to maximize accuracy, groups with increased risk factors should be targeted for standardized use of the pressure gauge.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Anesthesiologists
/
Intubation, Intratracheal
Type of study:
Clinical_trials
/
Observational_studies
/
Risk_factors_studies
Limits:
Humans
Language:
En
Journal:
Sci Rep
Year:
2023
Document type:
Article
Country of publication:
United kingdom