Your browser doesn't support javascript.
loading
Repeated Measurements Are Necessary for Evaluating Accurate Diurnal Rhythm Using a Self-Intraocular Pressure Measurement Device.
Shigemoto, Yumi; Hasebe, Yuka; Kitamura, Kazuyoshi; Fukuda, Yoshiko; Sakamoto, Masako; Matsubara, Mio; Minaguchi, Shinya; Kashiwagi, Kenji.
Affiliation
  • Shigemoto Y; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Hasebe Y; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Kitamura K; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Fukuda Y; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Sakamoto M; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Matsubara M; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Minaguchi S; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
  • Kashiwagi K; Department of Ophthalmology, Faculty of Medicine, University of Yamanashi, Yamanashi 409-3898, Japan.
J Clin Med ; 12(7)2023 Mar 23.
Article in En | MEDLINE | ID: mdl-37048542
ABSTRACT

PURPOSE:

To investigate how many tests need to be performed to adequately assess intraocular pressure (IOP) diurnal change using a self-measuring rebound tonometer among glaucoma patients. SUBJECTS AND

METHODS:

Adult patients with primary open-angle glaucoma were included. IOP was measured in the morning (6 AM to 9 AM), afternoon (12 PM to 3 PM), and at night (6 PM to 9 PM) for seven consecutive days. Twenty-four (7 males and 17 females, mean age 59.5 ± 11.0 years) patients who successfully measured IOP at least three times per day during the correct time periods for four days were subjected to analysis.

RESULTS:

The IOP rhythm was significantly greater on the first day of measurement (6.6 ± 3.6 mmHg) than that averaged during subsequent days (4.4 ± 2.2 mmHg). The time of the highest and lowest IOP measurements on the first day of IOP measurement and during the entire measurement period coincided in 72.9% and 64.6% of cases, respectively. The concordance rate of the highest IOP time between the whole measurement period and each measurement day was less than 60%.

CONCLUSION:

The diurnal IOP rhythm measured by the patients themselves was not consistent, and multiple days of measurements may be necessary to correctly assess diurnal IOP rhythm.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Clin Med Year: 2023 Document type: Article Affiliation country: Japan

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Clin Med Year: 2023 Document type: Article Affiliation country: Japan