Feasibility and reproducibility of semi-automated longitudinal strain analysis: a comparative study with conventional manual strain analysis.
Peng, Gui-Juan; Luo, Shu-Yu; Zhong, Xiao-Fang; Lin, Xiao-Xuan; Zheng, Ying-Qi; Xu, Jin-Feng; Liu, Ying-Ying; Chen, Li-Xin.
Affiliation
  • Peng GJ; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China.
  • Luo SY; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China.
  • Zhong XF; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China.
  • Lin XX; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China.
  • Zheng YQ; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China.
  • Xu JF; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China. jinfengxu@ext.jnu.edu.cn.
  • Liu YY; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China. yingyingliu@ext.jnu.edu.cn.
  • Chen LX; Shenzhen Medical Ultrasound Engineering Center, Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, 518020, China. 2116000950@qq.com.
Cardiovasc Ultrasound ; 21(1): 12, 2023 Jul 19.
Article in En | MEDLINE | ID: mdl-37464361
ABSTRACT

BACKGROUND:

Conventional approach to myocardial strain analysis relies on a software designed for the left ventricle (LV) which is complex and time-consuming and is not specific for right ventricular (RV) and left atrial (LA) assessment. This study compared this conventional manual approach to strain evaluation with a novel semi-automatic analysis of myocardial strain, which is also chamber-specific.

METHODS:

Two experienced observers used the AutoStrain software and manual QLab analysis to measure the LV, RV and LA strains in 152 healthy volunteers. Fifty cases were randomly selected for timing evaluation.

RESULTS:

No significant differences in LV global longitudinal strain (LVGLS) were observed between the two methods (-21.0% ± 2.5% vs. -20.8% ± 2.4%, p = 0.230). Conversely, RV longitudinal free wall strain (RVFWS) and LA longitudinal strain during the reservoir phase (LASr) measured by the semi-automatic software differed from the manual analysis (RVFWS -26.4% ± 4.8% vs. -31.3% ± 5.8%, p < 0.001; LAS 48.0% ± 10.0% vs. 37.6% ± 9.9%, p < 0.001). Bland-Altman analysis showed a mean error of 0.1%, 4.9%, and 10.5% for LVGLS, RVFWS, and LASr, respectively, with limits of agreement of -2.9,2.6%, -8.1,17.9%, and -12.3,33.3%, respectively. The semi-automatic method had a significantly shorter strain analysis time compared with the manual method.

CONCLUSIONS:

The novel semi-automatic strain analysis has the potential to improve efficiency in measurement of longitudinal myocardial strain. It shows good agreement with manual analysis for LV strain measurement.