Subclinical Keratoconus Detection and Characterization Using Motion-Tracking Brillouin Microscopy.
Randleman, J Bradley; Zhang, Hongyuan; Asroui, Lara; Tarib, Imane; Dupps, William J; Scarcelli, Giuliano.
Affiliation
  • Randleman JB; Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio; Cleveland Clinic Lerner College of Medicine of Case Western Reserve University, Cleveland, Ohio. Electronic address: randlej@ccf.org.
  • Zhang H; Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio.
  • Asroui L; Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio.
  • Tarib I; Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio.
  • Dupps WJ; Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio; Cleveland Clinic Lerner College of Medicine of Case Western Reserve University, Cleveland, Ohio; Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio.
  • Scarcelli G; Fischell Department of Bioengineering, University of Maryland, College Park, Maryland.
Ophthalmology ; 131(3): 310-321, 2024 Mar.
Article in En | MEDLINE | ID: mdl-37839561
ABSTRACT

PURPOSE:

To characterize focal biomechanical alterations in subclinical keratoconus (SKC) using motion-tracking (MT) Brillouin microscopy and evaluate the ability of MT Brillouin metrics to differentiate eyes with SKC from normal control eyes.

DESIGN:

Prospective cross-sectional study.

PARTICIPANTS:

Thirty eyes from 30 patients were evaluated, including 15 eyes from 15 bilaterally normal patients and 15 eyes with SKC from 15 patients.

METHODS:

All patients underwent Scheimpflug tomography and MT Brillouin microscopy using a custom-built device. Mean and minimum MT Brillouin values within the anterior plateau region and anterior 150 µm were generated. Scheimpflug metrics evaluated included inferior-superior (IS) value, maximum keratometry (Kmax), thinnest corneal thickness, asymmetry indices, Belin/Ambrosio display total deviation, and Ambrosio relational thickness. Receiver operating characteristic (ROC) curves were generated for all Scheimpflug and MT Brillouin metrics evaluated to determine the area under the ROC curve (AUC), sensitivity, and specificity for each variable. MAIN OUTCOME

MEASURES:

Discriminative performance based on AUC, sensitivity, and specificity.

RESULTS:

No significant differences were found between groups for age, sex, manifest refraction spherical equivalent, corrected distance visual acuity, Kmax, or KISA% index. Among Scheimpflug metrics, significant differences were found between groups for thinnest corneal thickness (556 µm vs. 522 µm; P < 0.001), IS value (0.29 diopter [D] vs. 1.05 D; P < 0.001), index of vertical asymmetry (IVA; 0.10 vs. 0.19; P < 0.001), and keratoconus index (1.01 vs. 1.05; P < 0.001), and no significant differences were found for any other Scheimpflug metric. Among MT Brillouin metrics, clear differences were found between control eyes and eyes with SKC for mean plateau (5.71 GHz vs. 5.68 GHz; P < 0.0001), minimum plateau (5.69 GHz vs. 5.65 GHz; P < 0.0001), mean anterior 150 µm (5.72 GHz vs. 5.68 GHz; P < 0.0001), and minimum anterior 150 µm (5.70 GHz vs. 5.66 GHz; P < 0.001). All MT Brillouin plateau and anterior 150 µm mean and minimum metrics fully differentiated groups (AUC, 1.0 for each), whereas the best performing Scheimpflug metrics were keratoconus index (AUC, 0.91), IS value (AUC, 0.89), and IVA (AUC, 0.88).

CONCLUSIONS:

Motion-tracking Brillouin microscopy metrics effectively characterize focal corneal biomechanical alterations in eyes with SKC and clearly differentiated these eyes from control eyes, including eyes that were not differentiated accurately using Scheimpflug metrics. FINANCIAL DISCLOSURE(S) Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.