Effectiveness of low tube voltage scan in the exposure dose for lenses during paediatric thoracic CT examination: anthropomorphic phantoms study.
Masuda, Takanori; Kiguchi, Masao; Fujioka, Chikako; Oku, Takayuki; Ishibashi, Toru; Katsunuma, Yasushi; Yoshitake, Takayasu; Abe, Shuji; Awai, Kazuo.
Affiliation
  • Masuda T; Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, 288, Matsushima, Kurashiki, Okayama 701-0193, Japan.
  • Kiguchi M; Department of Radiology, Hiroshima University, 2-3, Kasumi, Minami-ku, Hiroshima 734-0037, Japan.
  • Fujioka C; Department of Radiology, Hiroshima University, 2-3, Kasumi, Minami-ku, Hiroshima 734-0037, Japan.
  • Oku T; Department of Radiological Technologist, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima 730-8655, Japan.
  • Ishibashi T; Department of Radiological Technologist, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima 730-8655, Japan.
  • Katsunuma Y; Department of Radiological Technology, Tokai University Oiso Hospital, 143, Shimokasuya, Iseharashi, Kanagawa 259-1193, Japan.
  • Yoshitake T; TOWCAR WORKS Co., Ltd Hoashi 233-1, Kusu-machi, Kusugun, Oita 879-4403, Japan.
  • Abe S; Department of Radiological Technologist, Osaka College of High Technology, 1-2-43, Miyahara, Yogogawa-ku, Osaka 532-0003, Japan.
  • Awai K; Department of Diagnostic Radiology, Hiroshima University, 2-3, Kasumi, Minami-ku, Hiroshima 734-0037, Japan.
Radiat Prot Dosimetry ; 200(2): 143-148, 2024 Feb 16.
Article in En | MEDLINE | ID: mdl-37987195
ABSTRACT
To determine whether using lower-tube voltage reduces the scattered dose for the lens during paediatric thoracic computed tomography (CT). Two paediatric anthropomorphic phantoms (ATOM Phantom, CIRS, Norfolk, Virginia, USA) representing a newborn and 5-year-old were placed on the gantry of CT scanner, and optically stimulated luminescence dosemeters were placed on the left and right lenses, in front of the left and right thyroid glands, in front of the left and right mammary glands, and in front of and behind the mammary gland level and we measured scattered dose of the optically stimulated luminescence dosemeter was compared for each phantom between 80 and 120 kVp. Significant differences were observed in the scatter doses for the lens between 80 and 120 kVp (p < 0.01). Compared with the 120 kVp scan, the scatter doses for the lens were ~15-40% lower in newborn and 5-year-olds using the 80 kVp scan during paediatric CT.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tomography, X-Ray Computed / Lens, Crystalline Limits: Child / Child, preschool / Humans / Newborn Language: En Journal: Radiat Prot Dosimetry Year: 2024 Document type: Article Affiliation country: Japan

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tomography, X-Ray Computed / Lens, Crystalline Limits: Child / Child, preschool / Humans / Newborn Language: En Journal: Radiat Prot Dosimetry Year: 2024 Document type: Article Affiliation country: Japan