Validation, performance, and reliability of two automated tests for vitamin B12 and folate assay.
Bianciardi, Simone; Tesi, Giulia; Cerutti, Helena; Cartocci, Alessandra; Guerranti, Roberto; Silvestrini, Caterina; Gori, Sabrina; Bandini, Tommaso; Brogi, Alessandra; Leoncini, Roberto.
Affiliation
  • Bianciardi S; DIESSE Diagnostica Senese S.p.A. Società Benefit, Monteriggioni, Siena, Italy.
  • Tesi G; DIESSE Diagnostica Senese S.p.A. Società Benefit, Monteriggioni, Siena, Italy.
  • Cerutti H; DIESSE Diagnostica Senese S.p.A. Società Benefit, Monteriggioni, Siena, Italy.
  • Cartocci A; Department of Medical Biotechnology, University of Siena, Siena, Italy.
  • Guerranti R; Laboratorio Patologia Clinica, Azienda Ospedaliero-Universitaria Senese, Siena, Italy.
  • Silvestrini C; Laboratorio Patologia Clinica, Azienda Ospedaliero-Universitaria Senese, Siena, Italy.
  • Gori S; Laboratorio Patologia Clinica, Azienda Ospedaliero-Universitaria Senese, Siena, Italy.
  • Bandini T; DIESSE Diagnostica Senese S.p.A. Società Benefit, Monteriggioni, Siena, Italy.
  • Brogi A; DIESSE Diagnostica Senese S.p.A. Società Benefit, Monteriggioni, Siena, Italy.
  • Leoncini R; Laboratorio Patologia Clinica, Azienda Ospedaliero-Universitaria Senese, Siena, Italy.
J Public Health Res ; 12(4): 22799036231217803, 2023 Oct.
Article in En | MEDLINE | ID: mdl-38108046
ABSTRACT

Background:

Deficiency of Vitamin B12 and folate may determine hematological, neurological, and metabolic alterations; therefore, an accurate quantification of their serum levels is required, especially in the presence of symptoms that might suggest a deficiency. CHORUS VIT B12 and CHORUS FOLATE are two automated immunoassays, developed to quantify vitamin B12 and folate, respectively, in human serum. Design and

methods:

This single-center, non-pharmacological, diagnostic study described the validation and characterization of CHORUS VIT B12 and CHORUS FOLATE, with a specific focus on performance, precision, and reliability. For each assay, 500 serum samples were analyzed. A comparison between CHORUS assays and commercially available kit was also performed.

Results:

For CHORUS VIT B12 the lower limit of quantification (LLoQ) was 165.0 pg/mL and the upper LoQ (ULoQ) was 1846.8 pg/mL. The assay was linear within the calibration range (150-2000 pg/mL) and the accuracy was described with the International Standard Vitamin B12, Serum Folate, HOLO TC (NIBSC code 03/178), with a mean recovery on two lots of 111%. For CHORUS FOLATE (calibration range of 2.0-20.0 ng/mL), LLoQ was 2.0 ng/mL and ULoQ 19.6 ng/mL. The linearity was demonstrated from 2.4 to 20.0 ng/mL; the accuracy was described with the International Standard mentioned above, achieving a mean recovery on three lots of 92%. The lowest and highest values of both CHORUS and COBAS kits were similar and the median values did not significantly vary.

Conclusion:

CHORUS VIT B12 and CHORUS FOLATE performed well, accurately, and reliably in quantifying vitamin B12 and folate in human serum.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Public Health Res Year: 2023 Document type: Article Affiliation country: Italy

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Public Health Res Year: 2023 Document type: Article Affiliation country: Italy