Finite element modeling to predict the influence of anatomic variation and implant placement on performance of biological intervertebral disc implants.
Koga, Maho; Kim, Byumsu; Lintz, Marianne; Kirnaz, Sertaç; Goldberg, Jacob L; Hussain, Ibrahim; Medary, Branden; Meyers, Kathleen N; Maher, Suzanne A; Härtl, Roger; Bonassar, Lawrence J.
Affiliation
  • Koga M; Meinig School of Biomedical Engineering Cornell University Ithaca New York USA.
  • Kim B; Sibley School of Mechanical and Aerospace Engineering Cornell University Ithaca New York USA.
  • Lintz M; Meinig School of Biomedical Engineering Cornell University Ithaca New York USA.
  • Kirnaz S; Weill Cornell Medicine New York New York USA.
  • Goldberg JL; Weill Cornell Medicine New York New York USA.
  • Hussain I; Weill Cornell Medicine New York New York USA.
  • Medary B; Weill Cornell Medicine New York New York USA.
  • Meyers KN; Hospital for Special Surgery New York New York USA.
  • Maher SA; Hospital for Special Surgery New York New York USA.
  • Härtl R; Weill Cornell Medicine New York New York USA.
  • Bonassar LJ; Meinig School of Biomedical Engineering Cornell University Ithaca New York USA.
JOR Spine ; 6(4): e1307, 2023 Dec.
Article in En | MEDLINE | ID: mdl-38156058
ABSTRACT

Background:

Tissue-engineered intervertebral disc (TE-IVD) constructs are an attractive therapy for treating degenerative disc disease and have previously been investigated in vivo in both large and small animal models. The mechanical environment of the spine is notably challenging, in part due to its complex anatomy, and implants may require additional mechanical support to avoid failure in the early stages of implantation. As such, the design of suitable support implants requires rigorous validation.

Methods:

We created a FE model to simulate the behavior of the IVD cages under compression specific to the anatomy of the porcine cervical spine, validated the FE model using an animal model, and predicted the effects of implant location and vertebral angle of the motion segment on implant behavior. Specifically, we tested anatomical positioning of the superior vertebra and placement of the implant. We analyzed corresponding stress and strain distributions.

Results:

Results demonstrated that the anatomical geometry of the porcine cervical spine led to concentrated stress and strain on the posterior side of the cage. This stress concentration was associated with the location of failure of the cages reported in vivo, despite superior mechanical properties of the implant. Furthermore, placement of the cage was found to have profound effects on migration, while the angle of the superior vertebra affected stress concentration of the cage.

Conclusions:

This model can be utilized both to inform surgical procedures and provide insight on future cage designs and can be adopted to models without the use of in vivo animal models.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: JOR Spine / JOR spine Year: 2023 Document type: Article Country of publication: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: JOR Spine / JOR spine Year: 2023 Document type: Article Country of publication: United States