Reconstructive interpolation for pulse wave estimation to improve local PWV measurement of carotid artery.
Gu, Ouyang; He, Bingbing; Xiong, Li; Zhang, Yufeng; Li, Zhiyao; Lang, Xun.
Affiliation
  • Gu O; Department of Electronic Engineering, Information School, Yunnan University, Kunming, 650091, Yunnan, China.
  • He B; Department of Electronic Engineering, Information School, Yunnan University, Kunming, 650091, Yunnan, China. hebingbing@ynu.edu.cn.
  • Xiong L; Department of Electronic Engineering, Information School, Yunnan University, Kunming, 650091, Yunnan, China.
  • Zhang Y; Department of Electronic Engineering, Information School, Yunnan University, Kunming, 650091, Yunnan, China.
  • Li Z; Department of Ultrasound, the Third Affiliated Hospital of Kunming Medical College, Kunming, 650118, China.
  • Lang X; Department of Electronic Engineering, Information School, Yunnan University, Kunming, 650091, Yunnan, China.
Med Biol Eng Comput ; 62(5): 1459-1473, 2024 May.
Article in En | MEDLINE | ID: mdl-38252371
ABSTRACT
Ultrasonic transit time (TT)-based local pulse wave velocity (PWV) measurement is defined as the distance between two beam positions on a segment of common carotid artery (CCA) divided by the TT in the pulse wave propagation. However, the arterial wall motions (AWMs) estimated from ultrasonic radio frequency (RF) signals with a limited number of frames using the motion tracking are typically discrete. In this work, we develop a method involving motion tracking combined with reconstructive interpolation (MTRI) to reduce the quantification errors in the estimated PWs, and thereby improve the accuracy of the TT-based local PWV measurement for CCA. For each beam position, normalized cross-correlation functions (NCCFs) between the reference (the first frame) and comparison (the remaining frames) RF signals are calculated. Thereafter, the reconstructive interpolation is performed in the neighborhood of the NCCFs' peak to identify the interpolation-deduced peak locations, which are more exact than the original ones. According to which, the improved AWMs are obtained to calculate their TT along a segment of the CCA. Finally, the local PWV is measured by applying a linear regression fit to the time-distance result. In ultrasound simulations based on the pulse wave propagation models of young, middle-aged, and elderly groups, the MTRI method with different numbers of interpolated samples was used to estimate AWMs and local PWVs. Normalized root mean squared errors (NRMSEs) between the estimated and preset values of the AWMs and local PWVs were calculated and compared with ones without interpolation. The means of the NRMSEs for the AWMs and local PWVs based on the MTRI method with one interpolated sample decrease from 1.14% to 0.60% and 7.48% to 4.61%, respectively. Moreover, Bland-Altman analysis and coefficient of variation were used to validate the performance of the MTRI method based on the measured local PWVs of 30 healthy subjects. In conclusion, the reconstructive interpolation for the pulse wave estimation improves the accuracy and repeatability of the carotid local PWV measurement.