Your browser doesn't support javascript.
loading
Effects of sterilization, conditioning, and thermal aging on the retention of zirconia hybrid abutments: A laboratory study.
Alghanaim, Nalah; Kern, Matthias; Yazigi, Christine; Chaar, M Sad.
Affiliation
  • Alghanaim N; Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry, Christian-Albrechts University, Kiel, Germany.
  • Kern M; Department of Prosthodontics, Faculty of Dentistry, King Abdulaziz University, Jeddah, Saudi Arabia.
  • Yazigi C; Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry, Christian-Albrechts University, Kiel, Germany.
  • Chaar MS; Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry, Christian-Albrechts University, Kiel, Germany.
Clin Implant Dent Relat Res ; 26(4): 679-687, 2024 Aug.
Article in En | MEDLINE | ID: mdl-38680076
ABSTRACT

OBJECTIVE:

To investigate the effects of sterilization, conditioning method, and thermal aging on the retentive strength of two-piece zirconia abutments. MATERIALS AND

METHODS:

In total, 128 stock zirconia abutments were divided into four groups (n = 32) according to the conditioning parameters (A) air-abrasion using 50 µm alumina particles/1.0 bar, (B) 50 µm/2.0 bar, (C) 100 µm/1.0 bar, and (D) 100 µm/2.0 bar. All abutments were bonded onto titanium bases using DTK adhesive resin and stored in water bath (37°C) for 72 h. Each group was subdivided into two subgroups (n = 16), group 1 was disinfected, whereas group 2 followed disinfection and autoclave sterilization. Half of the specimens of each subgroup (n = 8) was directly subjected to the axial retention test (groups N), while the other half was first subjected to 150 days of thermocycling followed by retention test (groups T). Statistical analysis was performed with three-way ANOVA, additional statistical analysis was performed by using separate one-way ANOVAs followed by the Tukey's post-hoc test for post hoc pairwise comparisons among groups.

RESULTS:

The highest median retention strength was recorded for group B2N (1390 N), whereas the lowest strength was recorded for group C1T (688 N). No significant interaction (p ≥ 0.05) was detected between the different variables; conditioning method, sterilization, and the thermal cycling regarding the effect on the resulting retention. However, the sterilization always showed a positive effect. Thermocycling presented an adversely significant effect only in the absence of sterilization (p < 0.05), with the exception of subgroups A. For the sterilized groups, thermocycling had no statistically significant effect on the retention.

CONCLUSION:

Steam autoclaving increased the retention of hybrid zirconia abutments. DTK adhesive resin for two-piece zirconia abutments performed well after sterilization and thermocycling.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Zirconium / Sterilization / Dental Abutments Limits: Humans Language: En Journal: Clin Implant Dent Relat Res Journal subject: ODONTOLOGIA Year: 2024 Document type: Article Affiliation country: Germany

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Zirconium / Sterilization / Dental Abutments Limits: Humans Language: En Journal: Clin Implant Dent Relat Res Journal subject: ODONTOLOGIA Year: 2024 Document type: Article Affiliation country: Germany