Endocardial lead insulation wear in a scanning and optical microscope.
Ulman, Mateusz; Boczar, Krzysztof; Holcman, Katarzyna; Ziabka, Magdalena; Debski, Maciej; Lelakowski, Jacek; Zabek, Andrzej.
Affiliation
  • Ulman M; Department of Electrocardiology, The John Paul II Hospital, Cracow, Poland.
  • Boczar K; Department of Electrocardiology, The John Paul II Hospital, Cracow, Poland.
  • Holcman K; Department of Cardiac and Vascular Diseases, The John Paul II Hospital, Cracow, Poland.
  • Ziabka M; Institute of Cardiology, Jagiellonian University Medical College, Cracow, Poland.
  • Debski M; Department of Ceramics and Refractories, Faculty of Materials Science and Ceramics, AGH University, Cracow, Poland.
  • Lelakowski J; Norwich Medical School, University of East Anglia, UK.
  • Zabek A; Department of Electrocardiology, The John Paul II Hospital, Cracow, Poland.
Adv Clin Exp Med ; 2024 Jun 27.
Article in En | MEDLINE | ID: mdl-38934334
ABSTRACT

BACKGROUND:

The path and interaction of leads within the cardiovascular system are influenced by various factors, including the implantation technique. Furthermore, the multifaceted composition of these leads, often comprising multiple materials, can contribute to their potential degradation and wear over time.

OBJECTIVES:

Our aim was to investigate the wear of lead insulation following the removal of transvenous leads and pinpoint the regions of the lead most vulnerable to damage. MATERIAL AND

METHODS:

We undertook a prospective analysis of patients from a single tertiary center who underwent transvenous lead explantation (TLE) between October 1, 2013, and July 31, 2015. Specifically, our examination focused on endocardial leads removed using simple screw-out and gentle traction techniques. Subsequent lead evaluations were conducted utilizing scanning electron and optical microscopes.

RESULTS:

Among the 86 patients who underwent the TLE procedure, 26 patients (30%) required the removal of 39 leads through simple traction. Inspection using scanning electron microscopy consistently indicated insulation damage across all leads. A total of 347 damaged sites were identified 261 without lead unsealing and 86 exhibiting unsealing. Notably, the sections of the leads located within the intra-pocket area demonstrated the highest vulnerability to damage (odds ratio (OR) = 9.112, 95% confidence interval (95% CI) 3.326-24.960), whereas the intravenous regions displayed the lowest susceptibility (OR 0.323, 95% CI 0.151-0.694).

CONCLUSIONS:

Our study reveals that all evaluated leads exhibited insulation damage, with the intra-pocket segments manifesting a notably higher prevalence of damage than the intravenous segments.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Adv Clin Exp Med Year: 2024 Document type: Article Affiliation country: Poland

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Adv Clin Exp Med Year: 2024 Document type: Article Affiliation country: Poland
...