Your browser doesn't support javascript.
loading
A novel approach to calculate the required volume of air for bronchial blockers in young children.
Zhu, Change; Zhang, Saiji; Zhang, Mazhong; Wei, Rong.
Affiliation
  • Zhu C; Department of Anesthesiology, Shanghai Children's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Zhang S; Department of Anesthesiology, Shanghai Children's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Zhang M; Department of Anesthesiology, Shanghai Children's Medical Center, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Wei R; Department of Anesthesiology, Shanghai Children's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Paediatr Anaesth ; 2024 Jul 12.
Article in En | MEDLINE | ID: mdl-38994735
ABSTRACT

INTRODUCTION:

Bronchial blocker balloons inflated with small volumes of air increase balloon pressure, involving a risk of airway injury especially in young children. However, there are no established guidelines regarding the appropriate volumes of air required to provide safe bronchial occlusion.

METHODS:

This study aimed to introduce a novel method for calculating the amount of air required for safe bronchial blocker balloon occlusion for one lung anesthesia in young children. We included 79 pediatric patients who underwent video-assisted thoracoscopic surgery at our hospital. Preoperatively, the balloon pressure and corresponding diameter of 5F bronchial blockers inflated with different volumes of air were measured. Intraoperatively, bronchial diameters measured by computerized tomographic scans were matched to the ex vivo measured balloon diameters. The quality of lung isolation, incidence of balloon repositioning, and airway injury were documented. Postoperatively, airway injury was evaluated through fiberoptic bronchoscopy.

RESULTS:

Balloon pressure and balloon diameter showed linear and nonlinear correlations with volume, respectively. The median lengths of the right and left mainstem bronchi were median (interquartile range) range 5.3 mm (4.5-6.3) 2.7-8.15 and 21.8 (19.6-23.4) 14-29, respectively. Occluding the left mainstem bronchus required <1 mL of air, with a balloon pressure of 27 cm H2O. The isolation quality was high with no case of mucosal injury or displacement. Occluding the right mainstem bronchus required a median air volume of 1.3 mL, with a median balloon pressure of 44 cm H2O. One patient had poor lung isolation due to a tracheal bronchus and another developed mild and transient airway injury.

CONCLUSION:

The bronchial blocker cuff should be regarded as a high-pressure balloon. We introduced a new concept for safe bronchial blocker balloon occlusion for one-lung ventilation in small children.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Paediatr Anaesth Journal subject: ANESTESIOLOGIA / PEDIATRIA Year: 2024 Document type: Article Affiliation country: China

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Paediatr Anaesth Journal subject: ANESTESIOLOGIA / PEDIATRIA Year: 2024 Document type: Article Affiliation country: China
...