The Minimum 5-Year Follow up of a Highly Versatile Distally Anchored Femoral Revision System With Hydroxyapatite Coating.
Arthroplast Today
; 26: 101185, 2024 Apr.
Article
in En
| MEDLINE
| ID: mdl-39006859
ABSTRACT
Background:
Total hip arthroplasty is one of the most successful operations medicine can offer. As more patients undergo total hip arthroplasty the revision burden increases proportionately. This is a cohort study of the Arcos Modular Femoral Revision System.Methods:
The primary outcome was rerevision of the femoral component. Secondary outcomes include complications and radiological and clinical outcomes over 5 years.Results:
A total of 74 patients were recruited, and the femoral survival rate was 100% at 5 years with 12 censorships. A total of 31 patients were given Proprosky 3/4 preoperatively. Eleven patients underwent further procedures; however, no femoral components were revised. Kaplan-Meier analysis was performed by a biostatistician. Patients demonstrated a consistent and sustained improvement in Harris hip score, Oxford hip score, and EQ-5D. Radiological review revealed minimal and stable lysis around the femoral components. Conculsion The ARCOS Modular Femoral Revision System was designed to offer a range of options to allow femoral bone stock preservation and avoiding conversion to an endoprosthetic and seems to be effective in the medium term.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Language:
En
Journal:
Arthroplast Today
Year:
2024
Document type:
Article
Country of publication:
United States