Your browser doesn't support javascript.
loading
Respiratory-resolved five-dimensional flow cardiovascular magnetic resonance : In-vivo validation and respiratory-dependent flow changes in healthy volunteers and patients with congenital heart disease.
Weiss, Elizabeth K; Baraboo, Justin; Rigsby, Cynthia K; Robinson, Joshua D; Ma, Liliana; Falcão, Mariana B L; Roy, Christopher W; Stuber, Matthias; Markl, Michael.
Affiliation
  • Weiss EK; Department of Radiology, Northwestern University, Chicago, Illinois, USA; Department of Biomedical Engineering, Northwestern University, Evanston, Illinois, USA. Electronic address: elizabeth.weiss@northwestern.edu.
  • Baraboo J; Department of Radiology, Northwestern University, Chicago, Illinois, USA; Department of Biomedical Engineering, Northwestern University, Evanston, Illinois, USA.
  • Rigsby CK; Department of Radiology, Northwestern University, Chicago, Illinois, USA; Department of Medical Imaging, Ann & Robert Lurie Children's Hospital, Chicago, Illinois, USA.
  • Robinson JD; Department of Cardiology, Ann & Robert Lurie Children's Hospital, Chicago, Illinois, USA.
  • Ma L; Department of Radiology, Northwestern University, Chicago, Illinois, USA.
  • Falcão MBL; Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
  • Roy CW; Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
  • Stuber M; Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
  • Markl M; Department of Radiology, Northwestern University, Chicago, Illinois, USA; Department of Biomedical Engineering, Northwestern University, Evanston, Illinois, USA.
J Cardiovasc Magn Reson ; 26(2): 101077, 2024 Aug 02.
Article in En | MEDLINE | ID: mdl-39098573
ABSTRACT

BACKGROUND:

This study aimed to validate respiratory-resolved five-dimensional (5D) flow cardiovascular magnetic resonance (CMR) against real-time two-dimensional (2D) phase-contrast MRI, assess the impact of number of respiratory states, and measure the impact of respiration on hemodynamics in congenital heart disease (CHD) patients.

METHODS:

Respiratory-resolved 5D flow MRI-derived net and peak flow measurements were compared to real-time 2D phase-contrast MRI-derived measurements in 10 healthy volunteers. Pulmonary-to-systemic flow ratios (QpQs) were measured in 19 CHD patients and aortopulmonary collateral burden was measured in 5 Fontan patients. Additionally, the impact of number of respiratory states on measured respiratory-driven net flow changes was investigated in 10 healthy volunteers and 19 CHD patients (shunt physiology, n = 11, single ventricle disease [SVD], n = 8).

RESULTS:

There was good agreement between 5D flow MRI and real-time 2D phase-contrast-derived net and peak flow. Respiratory-driven changes had a good correlation (rho = 0.64, p < 0.001). In healthy volunteers, fewer than four respiratory states reduced measured respiratory-driven flow changes in veins (5.2 mL/cycle, p < 0.001) and arteries (1.7 mL/cycle, p = 0.05). Respiration drove substantial venous net flow changes in SVD (64% change) and shunt patients (57% change). Respiration had significantly greater impact in SVD patients compared to shunt patients in the right and left pulmonary arteries (46% vs 15%, p = 0.003 and 59% vs 20%, p = 0.002). QpQs varied by 37 ± 24% over respiration in SVD patients and 12 ± 20% in shunt patients. Aortopulmonary collateral burden varied by 118 ± 84% over respiration in Fontan patients. The smallest collateral burden was measured during active inspiration in all patients and the greatest burden was during active expiration in four of five patients. Reduced respiratory resolution blunted measured flow changes in the caval veins of shunt and SVD patients (p < 0.005).

CONCLUSIONS:

Respiratory-resolved 5D flow MRI measurements agree with real-time 2D phase contrast. Venous measurements are sensitive to number of respiratory states, whereas arterial measurements are more robust. Respiration has a substantial impact on caval vein flow, QpQs, and collateral burden in CHD patients.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Cardiovasc Magn Reson Journal subject: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Year: 2024 Document type: Article Country of publication: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Cardiovasc Magn Reson Journal subject: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Year: 2024 Document type: Article Country of publication: United kingdom