Your browser doesn't support javascript.
loading
Introduction and Validation of Low-Cost Ocular Biometry in Healthy and Cataractous Eyes Using a Thermally Tunable Swept-Source Laser.
Kendrisic, Milana; Georgiev, Stefan; Resch, Hemma; Steiner, Stefan; Salas, Matthias; Schmoll, Tilman; Drexler, Wolfgang; Findl, Oliver; Vass, Clemens; Leitgeb, Rainer A.
Affiliation
  • Kendrisic M; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria; Christian Doppler Laboratory for Innovative Optical Imaging and its Translation to Medicine, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.
  • Georgiev S; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria; VIROS, Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria.
  • Resch H; Department of Ophthalmology and Optometry, Vienna General Hospital, Vienna, Austria.
  • Steiner S; Department of Ophthalmology and Optometry, Vienna General Hospital, Vienna, Austria.
  • Salas M; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.
  • Schmoll T; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria; Carl Zeiss Meditec AG, Oberkochen, Germany.
  • Drexler W; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.
  • Findl O; VIROS, Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria.
  • Vass C; Department of Ophthalmology and Optometry, Vienna General Hospital, Vienna, Austria.
  • Leitgeb RA; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria; Christian Doppler Laboratory for Innovative Optical Imaging and its Translation to Medicine, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria. Ele
Am J Ophthalmol ; 269: 172-180, 2024 Aug 31.
Article in En | MEDLINE | ID: mdl-39218390
ABSTRACT

PURPOSE:

To introduce and validate a novel substantially lower-priced and rapid swept-source investigational optical biometer in healthy and cataractous eyes, using a thermally tuned laser diode used extensively in cell phones and data communication as an alternative swept source.

DESIGN:

Prospective accuracy, validity, and reliability analysis.

METHODS:

A total of 60 eyes of 59 subjects (29 eyes of 29 healthy subjects and 31 eyes of 30 cataract patients) were enrolled in a prospective comparative study at the Vienna General Hospital between August 2021 and April 2023. Averaged intraocular distances were acquired in 2.5 seconds from datasets consisting of 5000 consecutive A-scans at a single position by a low-cost swept-source optical biometry (SSOB) system. Instrument repeatability was assessed via standard deviations (SDs) and coefficients of variation (CoVs) of parameters such as axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and central corneal thickness (CCT). Healthy subjects and cataract patients were subsequently measured on the same day with the SSOB and a referential partial coherence interferometry (PCI) biometer (IOLMaster 500) to establish AL inter-device correlation (r) for instrument calibration. AL and ACD as shared parameters between both biometers were evaluated for their limits of agreements (LoA) using Bland-Altman analysis.

RESULTS:

Repeated measurements of AL, ACD, LT, and CCT revealed SDs of 18 µm, 12 µm, 12 µm, and 10 µm, respectively. All parameters except for CCT had a COV <1%. Except for 1 eye with white cataract, 59 eyes of 59 study participants with various degrees and types of cataract could be measured with both devices. The AL inter-device correlation was excellent (r > 0.99). The 95% LoAs between both biometers were -0.14 to 0.13 mm for AL and -0.28 to 0.25 mm for ACD.

CONCLUSIONS:

Optical biometry using a thermally tunable VCSEL swept-source light source has the potential to provide clinically relevant biometric parameters at an unprecedented 100-fold lower price point than currently used state-of-the-art optical biometers, paving the way for compact devices in remote care settings.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Am J Ophthalmol / Am. j. ophthalmol / American journal of ophthalmology Year: 2024 Document type: Article Affiliation country: Austria Country of publication: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Am J Ophthalmol / Am. j. ophthalmol / American journal of ophthalmology Year: 2024 Document type: Article Affiliation country: Austria Country of publication: United States