The effect of common parameters of bipolar stimulation on brain evoked potentials.
Clin Neurophysiol
; 167: 26-36, 2024 Aug 31.
Article
in En
| MEDLINE
| ID: mdl-39260137
ABSTRACT
OBJECTIVE:
To identify optimal bipolar stimulation parameters for robust generation of brain evoked potentials (BEPs), namely the interelectrode distance (IED) and the intensity of stimulation (IS), in cortical and axonal stimulation.METHODS:
In 15 patients who underwent awake surgery for brain tumor removal, BEPs were elicited at different values of IED and IS, respectively 5 mm-5 mA, 5 mm-10 mA, and 10 mm-10 mA. The number of BEPs elicited by stimulation, as well as the delays and amplitudes of the N1 waves were compared between the different groups of stimulation parameters and according to the stimulated brain structure (cortical vs. axonal).RESULTS:
The amplitudes of N1 increased with the intensity of bipolar stimulation, either in cortical or axonal stimulation, while N1 peak delays were not affected by the stimulation parameters. Furthermore, axonal stimulation produced more N1s than cortical stimulation, with lower latencies.CONCLUSIONS:
Understanding the relationship between stimulation parameters and BEP is of utmost importance to determine whether the generated N1 waves accurately reflect the underlying structural anatomy. Other factors, such as stimulation frequency or pulse width and shape, may also play a role and warrant further investigation.SIGNIFICANCE:
This study represents the first step in describing the influence of common bipolar stimulation parameters on robustness of BEPs by examining the impact of IED and IS on the N1 wave.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Language:
En
Journal:
Clin Neurophysiol
Journal subject:
NEUROLOGIA
/
PSICOFISIOLOGIA
Year:
2024
Document type:
Article
Affiliation country:
France
Country of publication:
Netherlands