ABSTRACT
The ellipsoid left heart assist device (E-LVAD) was implanted in eight patients suffering from intraoperative heart failure. It was not possible to remove these patients from extracorporeal circulation following an intracardiac procedure; therefore, implantation of the E-LVAD was performed during extracorporeal circulation. The inflow connector was pushed forward from a purse-string suture on the right superior pulmonary vein, across the mitral valve and into the left ventricle. The outflow connector was joined to the ascending aorta. In two patients, the artificial heart chamber was removed after complete recovery of the circulation; these patients, however, later died. In six other patients, untreatable right heart failure developed and these patients died with the pump in place. It is concluded, therefore, that the right heart must also be mechanically supported during postoperative heart failure.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Assisted Circulation / Heart, Artificial Limits: Adult / Aged / Female / Humans / Male / Middle aged Language: En Journal: Artif Organs Year: 1978 Document type: Article