Ontario accelerator dose intercomparison study.
J Can Assoc Radiol ; 32(1): 49-53, 1981 Mar.
Article in En | MEDLINE | ID: mdl-7217168
An ionometric dose intercomparison has been carried out on eight accelertors, with maximum photon energies from 6 to 32 meV, at five radiotherapy centres in Ontario. The ratio of the dose based on the clinically-employed rad/monitor unit to the dose measured by the committee representative had a mean value of 0.994, with a range of 7.3% and a coefficient of variation of 2.5%. The ratio of the dose measured by the institution's physicist to the dose measured by the committee representative had a mean value of 1.000 with a range of 6.9% and a coefficient of variation of 2.3%. Eight recommendations regarding dose calibration procedures are presented.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Radiotherapy Dosage Type of study: Guideline Limits: Humans Country/Region as subject: America do norte Language: En Journal: J Can Assoc Radiol Year: 1981 Document type: Article Country of publication: Canada