Characteristics of the ventilator pressure- and flow-trigger variables.
Sassoon, C S; Gruer, S E.
Affiliation
  • Sassoon CS; Pulmonary and Critical Care Section, UCI-Long Beach Medical Program, Veterans Affairs Medical Center, CA 90822, USA.
Intensive Care Med ; 21(2): 159-68, 1995 Feb.
Article in En | MEDLINE | ID: mdl-7775698
ABSTRACT
Pressure- and flow-triggering are available in the Puritan Bennett 7200ae and Siemens SV 300. Using a mechanical lung model, we described the characteristics of the pressure- and flow-triggered continuous positive airway pressure (CPAP) of both ventilators. In the Puritan Bennett 7200ae, the pressure-triggered CPAP is characterized by the relatively insufficient flow delivery after the triggering, resulting in a greater lung pressure-time product (total PTP) than the flow-triggered CPAP. Pressure support of 5 cmH2O results in total PTP less than that with flow-triggered CPAP. In the Siemens SV 300, total PTP with pressure- or flow-triggered CPAP is comparable. Total PTP is less with pressure- or flow-triggered CPAP of the Siemens SV 300 than that of the Puritan Bennett 7200ae, respectively. The application of small pressure- or flow-triggered pressure support in the Puritan Bennett 7200ae eliminates the difference. The impact of these differences on patient inspiratory muscle work remains to be determined.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Ventilators, Mechanical Type of study: Prognostic_studies Limits: Humans Language: En Journal: Intensive Care Med Year: 1995 Document type: Article Affiliation country: United States
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Ventilators, Mechanical Type of study: Prognostic_studies Limits: Humans Language: En Journal: Intensive Care Med Year: 1995 Document type: Article Affiliation country: United States