Your browser doesn't support javascript.
loading
Trace analysis in chiral separation of selected amino enantiomers.
Pronce, T; Tilquin, B.
Affiliation
  • Pronce T; Université Catholique de Louvain, Ecole de Pharmacie, Unité d'Analyse Chimique et Physico-Chimique des Médicaments (CHAM 72.30), Bruxelles, Belgium.
J Pharm Biomed Anal ; 14(8-10): 1175-84, 1996 Jun.
Article in En | MEDLINE | ID: mdl-8818031
A Crownpack CR(+) column was used for the highly sensitive quantitation of amino acid enantiomers. The use of a mobile phase of exclusively aqueous perchloric acid enabled the use of low-wavelength UV detection (200 nm) before on-line derivatization with o-phthaldialdehyde for fluorimetric detection. The use of simultaneous UV and fluorimetric detection permitted the simultaneous quantification of the impurity and its parent compound. The method was used for the evaluation of some commercially available amino acid standards. Enantiomeric impurities as low as 0.001% (10 ppm) can be determined in some cases. High precision for the determination of trace levels of D-amino acids in the presence of large amounts of corresponding L-enantiomer is demonstrated.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Amino Acids Type of study: Diagnostic_studies Language: En Journal: J Pharm Biomed Anal Year: 1996 Document type: Article Affiliation country: Belgium Country of publication: United kingdom
Search on Google