Prolonged prothrombin time and activated partial thromboplastin time due to underfilled specimen tubes with 109 mmol/L (3.2%) citrate anticoagulant.
Am J Clin Pathol
; 109(6): 754-7, 1998 Jun.
Article
in En
| MEDLINE
| ID: mdl-9620035
Underfilling of specimen tubes containing 129 mmol/L (3.8%) buffered citrate prolongs prothrombin time (PT) and activated partial thromboplastin time (APTT) values. We studied this phenomenon by using 109 mmol/L (3.2%) buffered citrate as the anticoagulant, anticipating some increase in tolerance to underfilling. Venous blood drawn from 12 healthy subjects and 30 patients receiving long-term oral warfarin therapy was mixed with 109 mmol/L buffered citrate solution in proportions equivalent to filling the collection tubes from 52% to 100% of capacity. Accurate PT values were obtained from normal specimens if the tubes were filled to 65% or more of capacity. Accurate PT results in the therapeutic range were obtained only with filling to 80% or more of capacity (using a "moderately sensitive" thromboplastin reagent, International Sensitivity Index [ISI] = 2.06) or 90% or more of capacity (using a "highly sensitive" thromboplastin reagent, ISI = 1.01). In contrast, APTT was much less tolerant to underfilling, with prolonged values observed in most specimens filled to less than 90% of capacity. No false low values were observed. Specimen tubes should be filled to at least 90% of capacity to avoid falsely elevated PT or APTT results, but values within the reference range may be acceptable even from underfilled tubes.
Search on Google
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Partial Thromboplastin Time
/
Prothrombin Time
/
Blood Specimen Collection
/
Citric Acid
/
Anticoagulants
Limits:
Adult
/
Female
/
Humans
/
Male
/
Middle aged
Language:
En
Journal:
Am J Clin Pathol
Year:
1998
Document type:
Article
Affiliation country:
United States
Country of publication:
United kingdom