Your browser doesn't support javascript.
loading
Comparative Study on Evaluating the Bladder Volume between BladderScan BVI9400 and Ultrasound System iU22 / 中国医疗器械杂志
Article in Zh | WPRIM | ID: wpr-265638
Responsible library: WPRO
ABSTRACT
<p><b>OBJECTIVE</b>To evaluate the accuracy of the latest BladderScan BVI9400 on measuring bladder volume.</p><p><b>METHODS</b>Two bladder phantoms were selected for investigating the accuracy of BVI9400. 341 patients with the iU22 ultrasound examinations were followed by BVI 9400. The difference and correlation between BVI9400 and iU22 were contrastively analyzed.</p><p><b>RESULTS</b>The relative difference between results from BVI9400 and phantom volume was 2.5% and 1.36%. There was a strong correlation for patients between BVI9400 and iU22 (R = 0.96, P < 0.001). The relative difference between BVI9400 and iU22 decreased with the increasing of bladder volume and had no significant difference with patient's gender (P > 0.1).</p><p><b>CONCLUSION</b>BladderScan BVI9400 had the ability of high accuracy and good stability of measured data. In view of quick and conveniences, BVI9400 could be as auxiliary equipment on pelvic tumor to evaluate whether the bladder volume during fractional radiotherapy was consistency with that during CT positioning.</p>
Subject(s)
Full text: 1 Database: WPRIM Main subject: Urinary Bladder / Diagnostic Imaging / Ultrasonography / Phantoms, Imaging / Methods Type of study: Diagnostic_studies Limits: Humans Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2015 Document type: Article
Full text: 1 Database: WPRIM Main subject: Urinary Bladder / Diagnostic Imaging / Ultrasonography / Phantoms, Imaging / Methods Type of study: Diagnostic_studies Limits: Humans Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2015 Document type: Article