Your browser doesn't support javascript.
loading
Determination of Aripiprazole in Human Plasma by HPLC / 中国药房
China Pharmacy ; (12)1991.
Article in Zh | WPRIM | ID: wpr-528870
Responsible library: WPRO
ABSTRACT

OBJECTIVE:

To establish a HPLC method for the determination of aripiprazole in human plasma by HPLC.

METHODS:

The plasma sample was determined by HPLC after subjected to liquid- liquid extraction .The separation was carried out on C18 column with clolumn temperature at 40℃. The mobile phase consisted of 0.03mol?L-1 ammonium acetate-acetonitrile(3466) with a flow rate of 0.8mL?min-1 and detective wavelength of 257nm.The sensitivity was 0.01 AUFS.

RESULTS:

The linear range of aripiprazole was 5.0~600.0ng?mL-1(r=0.999 5) .The recovery rate was above 90% .

CONCLUSION:

The method is sensitive,accurate and rapid,and suitable for the determination of aripiprazole in human plasma.
Key words
Full text: 1 Database: WPRIM Language: Zh Journal: China Pharmacy Year: 1991 Document type: Article
Full text: 1 Database: WPRIM Language: Zh Journal: China Pharmacy Year: 1991 Document type: Article