Clinical value evaluation of rapid detection method for the hematopoietic stem/progenitor cell of peripheral blood using the automated hematology analyzer / 中华检验医学杂志
Article in Zh | WPRIM | ID: wpr-871917
Responsible library: WPRO
ABSTRACT
Objective:To explore the performance and clinical application value of a rapid detection method for the hematopoietic stem/progenitor cell of peripheral blood using the automated hematology analyzer.Methods:Methodology validation and retrospective study. Collected sample from Fujian Medical University Union Hospital from January 2015 to December 2018, the peripheral blood of 4 patients with acute myeloid leukemia was first treated, and one healthy donor′s peripheral blood stem cell collection 5 times diluent, for the methodology validation. And the peripheral venous blood and 5-fold dilutions of peripheral blood stem cell collection, from 23 patients with hematopoietic stem cell transplantation and 22 healthy donors of allogeneic peripheral blood stem cell transplantation, used for consistency retrospective analysis. In the linear test, each of the peripheral blood and HPC collecting solutions from blood cell separator, which known CD34 +cell concentration, that was high-value samples for the expected upper limit (H) . Another low-value sample is normal saline (L) . According to the multiple proportion dilution, HPC was detected and regressed consistency test specimens were 126, EDTA-K 2 anticoagulant venous blood 78 and peripheral blood stem cell 48. Venous blood was collected at the same time, one tube of blood routine and HPC detection, the other tube flow cytometry (FCM) detection of CD34 +cells. Stem cell collection was diluted 5 times with sterilized saline and divided into two tubes. One tube was used to count whole blood cells and HPC, the other tube was used to detect CD34 +cells by FCM. The test results of the two instruments were compared, and the deviation was evaluated. Results:The background counting was 0×10 9/L and the carryover rate was 0.1%, conformed to the quality requirements of hematology analyzer, and the repeatability study imprecision ranged between 4.7%-18.8%. HPC of peripheral venous blood linear range (0-27.201×10 9/L), Stem cell collection was diluted 5 times linear range (0-0.878×10 9/L). The results of 126 samples detected by the hematology analyzer and FCM were compared. The correlation coefficient r2=0.960 1. When WBC>10×10 9/L, the results of the two instruments have a good consistency. The slope is between 0.95 and 1.05, and the relative bias is less than 30%. Conclusions:This study suggests that the hematology analyzer has a good linear range for detecting HPC, and has a good correlation with FCM. The hematology analyzer has the advantages of no pretreatment, convenient operation, a wide range of applications in detecting HPC specimens.
Full text: 1 Database: WPRIM Type of study: Diagnostic_studies / Observational_studies Language: Zh Journal: Chinese Journal of Laboratory Medicine Year: 2020 Document type: Article
Full text: 1 Database: WPRIM Type of study: Diagnostic_studies / Observational_studies Language: Zh Journal: Chinese Journal of Laboratory Medicine Year: 2020 Document type: Article