Determination of methylmercury in urine by direct mercury analyzer / 中华劳动卫生职业病杂志
Zhonghua laodong weisheng zhiyebing zazhi
; Zhonghua laodong weisheng zhiyebing zazhi;(12): 304-306, 2023.
Article
in Zh
| WPRIM
| ID: wpr-986033
Responsible library:
WPRO
ABSTRACT
Objective: To develop asolvent extraction-direct mercury analyzer method for determination of methylmercury in urine. Methods: After the urinehydrolyzesd by hydrobromic acid, methylmercury was extracted by tolueneand reverse-extracted from L-cysteine solution, it was then detectedbydirect mercuryanalyzer. Results: The linear range was 0.2-50.0 μg/L, and the related coefficient was 0.9999. The relative standard deviations (RSD) within the group were 5.04%-6.64%, and the RSD between the group were 5.65%-8.11 %. The average recovery efficiencies were 85.4%-95.5%. The detection limitation was 0.0482 μg/L and the quantification concentrations was 0.1607 μg/L. Conclusion: The method, which has low detection limit, high sensitivity, easy to operate, is stability for the determination of methylmercury in urine.
Key words
Full text:
1
Database:
WPRIM
Main subject:
Mercury
/
Methylmercury Compounds
Language:
Zh
Journal:
Zhonghua laodong weisheng zhiyebing zazhi
Year:
2023
Document type:
Article