Your browser doesn't support javascript.
loading
Application of ANSYS Fatigue Analysis in Type Selection and Evaluation of Bone Plate / 医用生物力学
Journal of Medical Biomechanics ; (6): E556-E560, 2023.
Article in Zh | WPRIM | ID: wpr-987985
Responsible library: WPRO
ABSTRACT
Objective By comparing the fatigue strength of type A and type B locking compression plates (LCP) in distal femoral plate, a theoretical evaluation method was provided for type selection of bone plate when testing its bending strength and fatigue performance. Methods Through bending strength performance test and fatigue performance test on bone plates with different types, combined with ANSYS Workbench, the finite element analysis on total deformation, von Mises stress and fatigue service life of bone plates were conducted. Results The fatigue strength of type A plate was 30.7% higher than that of type B plate, the stress of type A plate was lower than that of type B plate, and the minimum fatigue service life of type A plate was 17% higher than that of type B plate. Conclusions The fatigue performance of type A plate is better than that of type B plate, so the failure possibility of type A plate was lower than that of type B plate.The results provide references for assisting selection of different bone plates when testing the performance of two newly developed bone plates.
Key words
Full text: 1 Database: WPRIM Language: Zh Journal: Journal of Medical Biomechanics Year: 2023 Document type: Article
Full text: 1 Database: WPRIM Language: Zh Journal: Journal of Medical Biomechanics Year: 2023 Document type: Article