Different acquisition systems for heart rate variability analysis may lead to diverse outcomes
Oliveira Júnior, F.A. de; Pereira, R.A.; Silva, A.S.; Brito Alves, J.L. de; Costa-Silva, J.H.; Braga, V.A.; Balarini, C.M..
Afiliación
  • Oliveira Júnior, F.A. de; Universidade Federal da Paraíba. Centro de Ciências da Saúde. Departamento de Fisiologia e Patologia. João Pessoa. BR
  • Pereira, R.A.; Universidade Federal da Paraíba. Centro de Ciências da Saúde. Departamento de Educação Física. João Pessoa. BR
  • Silva, A.S.; Universidade Federal da Paraíba. Centro de Ciências da Saúde. Departamento de Educação Física. João Pessoa. BR
  • Brito Alves, J.L. de; Universidade Federal da Paraíba. Centro de Ciências da Saúde. Departamento de Nutrição. João Pessoa. BR
  • Costa-Silva, J.H.; Universidade Federal de Pernambuco. Departamento de Educação Física e Ciências do Esporte. Vitória de Santo Antão. BR
  • Braga, V.A.; Universidade Federal da Paraíba. Centro de Biotecnologia. João Pessoa. BR
  • Balarini, C.M.; Universidade Federal da Paraíba. Centro de Ciências da Saúde. Departamento de Fisiologia e Patologia. João Pessoa. BR
Braz. j. med. biol. res ; 55: e11720, 2022. tab, graf
Artículo en Inglés | LILACS-Express | LILACS | ID: biblio-1360239
Biblioteca responsable: BR1.1
ABSTRACT
Heart rate variability (HRV) is a relevant physiological variable for the estimation of cardiac autonomic function. Although the gold standard for HRV registration is the electrocardiogram (ECG), several applications (APPs) have been increasingly developed. The evaluation carried out by these devices must be compatible with ECG standards. The aim of this study was to compare the data obtained simultaneously with ECG and APP with chest heart rate transmitters. Fifty-six healthy individuals (28 men and 28 women) were evaluated at rest through a short simultaneous HRV measurement with both devices. Data from both acquisition systems were analyzed separately using their own analysis software and exported and analyzed using a validated software. Signal recordings were compatible between the two acquisition systems (Pearson r=0.99; P<0.0001). Although a high correlation was found for the HRV variables obtained in the time domain (Spearman r=0.99; P<0.0001), the correlation decreased in the frequency domain (Pearson r=0.85; P<0.0001) when two software programs were used. Comparison of the averages of spectral analysis parameters also showed differences when HRV data were analyzed separately in each device for low-frequency (LF) and high-frequency (HF) bands. Although the portability of these mobile devices allows for optimal HRV evaluation, the direct analysis obtained from these devices must be carefully evaluated with respect to frequency domain parameters.


Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: LILACS Idioma: Inglés Revista: Braz. j. med. biol. res Asunto de la revista: Biologia / Medicina Año: 2022 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade Federal da Paraíba/BR / Universidade Federal de Pernambuco/BR

Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: LILACS Idioma: Inglés Revista: Braz. j. med. biol. res Asunto de la revista: Biologia / Medicina Año: 2022 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade Federal da Paraíba/BR / Universidade Federal de Pernambuco/BR
...