Validez de un método fotométrico de medición de partículas en aire para el control de quirófanos y otras salas de ambiente controlado en los hospitales / Validity of a photometric method of counting particles in air for control of operating rooms and clean rooms in hospitals
Trauma (Majadahonda)
; 24(3): 195-199, jul.-sept. 2013. tab, ilus
Artículo
en Español
| IBECS
| ID: ibc-115582
Biblioteca responsable:
ES1.1
Ubicación: BNCS
RESUMEN
Objetivo:
Validar un sistema de medici¨®n fotom¨¦trica de part¨ªculas en aire para realizar los controles de bioseguridad ambiental de quir¨®fanos y otras salas de ambiente controlado en los hospitales. Material y m¨¦todos Se realizaron un total de 144 mediciones pareadas de aire, 88 en ocho quir¨®fanos con sistemas de ventilaci¨®n convencional y tres niveles de filtraci¨®n y 56 en nueve habitaciones sin instalaciones de ventilaci¨®n. Se midieron part¨ªculas de entre 0,1 a 10 ¦Ìm en mg/m3 por m¨¦todo fotom¨¦trico. Simult¨¢neamente, en cada sala se realiz¨® impacto de aire en placas de Petri con medio de cultivo mediante aparato aspirador de cabezal perforado, obteniendo tras cinco d¨ªas de incubaci¨®n recuento de unidades formadoras de colonias (UFC)/1000L. Finalmente se realiz¨® el c¨¢lculo de los puntos de corte ¨®ptimos para la medici¨®n de part¨ªculas mediante Curvas ROC para discriminar tres niveles de contaminaci¨®n.Resultados:
El coeficiente de correlaci¨®n entre las dos mediciones fue 0,779 (p<0,001). Los puntos de corte fueron para 0 UFC. 0,015, para >=10 UFC, 0,037, y para >=100 UFC, 0,053. La sensibilidad para estos puntos fue de 95,83%,100% y 100%, y los ¨ªndices Kappa registraron 0,51, 0,88 y 0,75 respectivamente. Conclusi¨®n En nuestro estudio la correlaci¨®n entre las medidas es notable; la medici¨®n fotom¨¦trica de part¨ªculas puede ser una alternativa para llevar a cabo alguno de los controles rutinarios de bioseguridad (AU)ABSTRACT
Objective:
Validate a photometric on-air particle measurement system in order to perform ambient bio-security checks in operating rooms and other clean rooms within hospitals. Material andmethods:
A total of 144 coupled air samples were performed, 88 in 8 different operating rooms with conventional ventilation systems and 3-level filtering and 56 in 9 different rooms without specific ventilation systems. Particles were measured in size between 0,1 to 10 ¦Ìm with the photometric system in mg/m3. Simultaneously in the rooms are performed the sampling air method with impact on culture media. Incubation period is 5 days and count is performed in colony-forming units (CFU/1000L). Finally, optimal cut-points in particles measure were calculated using ROC curves in order to discriminate the three different contamination levels.Results:
Spearma´s correlation coefficient was 0,779 p<0,001. Sensitivity for cut-points in particle counting for the different contamination levels was 95,83% ,100% and 100% respectively, and the corresponding Kappa indexes were 0,51; 0,88 and 0,75.Conclusion:
Correlation among measurements is notable. Particle measurement can be a worthy alternative for some of the routine bio-security checks (AU)
Buscar en Google
Colección:
Bases de datos nacionales
/
España
Contexto en salud:
ODS3 - Meta 3D Fortalecer la capacidad de alerta temprana, reducción y gestión de riesgos sanitarios nacionales y globales
/
Agenda de Salud Sostenible para las Américas
/
ODS3 - Salud y Bienestar
Problema de salud:
Riesgos Biológicos
/
Objetivo 6: Sistemas de información para la salud
/
Meta 3.9: Reducir la cantidad de muertes producidas por químicos peligrosos y la contaminación del aire, agua y suelo
Base de datos:
IBECS
Asunto principal:
Quirófanos
/
Fotometría
/
Sistemas de Información en Quirófanos
/
Contención de Riesgos Biológicos
Límite:
Femenino
/
Humanos
/
Masculino
Idioma:
Español
Revista:
Trauma (Majadahonda)
Año:
2013
Tipo del documento:
Artículo
Institución/País de afiliación:
Complexo Hospitalario Universitario de Ourense/España