Control de presiones venosa y arterial para evitar la disfunción del acceso vascular y su influencia en la dosis de diálisis / Venous and arterial pressure control to avoid vascular access dysfunction and his influence in the dyalisis dose
Sánchez Tocino, M Luz; Villoria González, Silvia; Muñoz García, Belén; Sánchez Martín, Alberto.
Afiliación
  • Sánchez Tocino, M Luz; Fundación Renal Íñigo Álvarez de Toledo (FRIAT). Salamanca. España
  • Villoria González, Silvia; Fundación Renal Íñigo Álvarez de Toledo (FRIAT). Salamanca. España
  • Muñoz García, Belén; Fundación Renal Íñigo Álvarez de Toledo (FRIAT). Salamanca. España
  • Sánchez Martín, Alberto; Fundación Renal Íñigo Álvarez de Toledo (FRIAT). Salamanca. España
Enferm. nefrol ; 19(3): 281-287, jul.-sept. 2016. tab, graf
Artículo en Español | IBECS | ID: ibc-156667
Biblioteca responsable: ES1.1
Ubicación: BNCS
RESUMEN

Introducción:

Las guías clínicas del acceso vascular, recomiendan el seguimiento y monitorización de la fistula arteriovenosa y recientes estudios el mantenimiento de la presión venosa y presión arterial dentro de unos límites seguros.

Objetivo:

Analizar las pautas de monitorización de diálisis en nuestros pacientes y ajustarlas para conseguir presiones dinámicas seguras. Valorar como afecta a la dosis de diálisis y realizar las modificaciones necesarias para recuperar esta de forma individualizada. Material y

métodos:

Se recogen datos de 29 pacientes, en tres periodos de tiempo de 15 sesiones; Periodo 0 pauta de diálisis habitual; Periodo 1 Flujo de bomba protector no modificable; Periodo 2 Flujo de bomba según presiones diarias y se modifica la pauta de diálisis. Se analizan las variables; flujo de bomba, presiones venosa, arterial y dosis de diálisis medido a través del Kt.

Resultados:

En el periodo 0, el 83% de los pacientes presenta una presión venosa >160 mm Hg y 21% presión arterial < -200 mm Hg. En el periodo 1, tras la intervención sobre el flujo, 20% de pacientes tiene presión venosa > de 160 mm Hg y 3,4% arterial < -200 mm Hg. En el periodo 3 el 100% pacientes están con presiones seguras. En relación a Kt periodo 0 media de 49l, periodo 1 media 46,8l y se recupera periodo 2 a 50,65l (p<0,001). Para ello en el periodo 2 se realizaron 21 intervenciones a 18 pacientes de forma individualizada.

Conclusión:

Se puede dializar siguiendo las últimas recomendaciones de límites de presiones dinámicas manteniendo la dosis de diálisis y actuando sobre otros parámetros de la pauta de manera individualizada (AU)
ABSTRACT

Introduction:

Clinical guidelines vascular access, recommend tracking and monitoring arteriovenous fistula and recent studies maintaining maintaining venous and arterial pressure in safe limits.

Objective:

To analyse dialysis monitoring patterns in our patients and to adjust them in order to get safe dynamic pressures. To value the influence in the dialysis dose and to carry out the required changes to restore it in an individualized way. Materials and

methods:

Dates of three periods in 29 patients has been taken in. Period zero Usual dialysis pattern Period one Protective blood flow not changeable. Period two Blood flow according to diary pressure and modified dialysis pattern. The variables analyzed are; blood flow, venous and arterial pressure, dialysis dose according to the Kt.

Results:

In period zero, 83% of the patients present a venous pressure >160 mm Hg and the 21% an arterial pressure < -200 mm Hg. In period one after the action above the flow 20% of the patients has a venous pressure > de 160 mm Hg and 3.4% an arterial pressure < -200 mm Hg. In period three, the 100% of the patients have safe pressures. In relation with the Kt, it was mean of 49l in the period zero, 46,8l in the period one and it regain in the period two to 50,65l (p<0,001). For that in the period two, 21 individualized interventions in 18 patients were performed.

Conclusions:

It is possible to dialyze according to the last guidelines in dynamic pressure limits, keeping the dialysis dose and acting in others parameters of the individualized pattern (AU)
Asunto(s)

Texto completo: Disponible Colección: Bases de datos nacionales / España Base de datos: IBECS Asunto principal: Presión Venosa / Monitores de Presión Sanguínea / Diálisis Renal / Monitoreo Ambulatorio de la Presión Arterial / Diálisis / Presión Arterial Tipo de estudio: Guía de práctica clínica / Estudio observacional Límite: Femenino / Humanos / Masculino Idioma: Español Revista: Enferm. nefrol Año: 2016 Tipo del documento: Artículo Institución/País de afiliación: Fundación Renal Íñigo Álvarez de Toledo (FRIAT)/España

Texto completo: Disponible Colección: Bases de datos nacionales / España Base de datos: IBECS Asunto principal: Presión Venosa / Monitores de Presión Sanguínea / Diálisis Renal / Monitoreo Ambulatorio de la Presión Arterial / Diálisis / Presión Arterial Tipo de estudio: Guía de práctica clínica / Estudio observacional Límite: Femenino / Humanos / Masculino Idioma: Español Revista: Enferm. nefrol Año: 2016 Tipo del documento: Artículo Institución/País de afiliación: Fundación Renal Íñigo Álvarez de Toledo (FRIAT)/España
...