Influence of short implants geometry on primary stability
González-Serrano, José; Molinero-Mourelle, Pedro; Pardal-Peláez, Beatriz; Sáez-Alcaide, Luis-Miguel; Ortega, Ricardo; López-Quiles, Juan.
Afiliación
  • González-Serrano, José; Complutense University. School of Dentistry. Department of Dental Clinical Specialties. Madrid. Spain
  • Molinero-Mourelle, Pedro; Complutense University. School of Dentistry. Department of Dental Clinical Specialties. Madrid. Spain
  • Pardal-Peláez, Beatriz; Complutense University. School of Dentistry. Department of Dental Clinical Specialties. Madrid. Spain
  • Sáez-Alcaide, Luis-Miguel; Complutense University. School of Dentistry. Department of Dental Clinical Specialties. Madrid. Spain
  • Ortega, Ricardo; Complutense University. School of Dentistry. Department of Dental Clinical Specialties. Madrid. Spain
  • López-Quiles, Juan; Complutense University. School of Dentistry. Department of Dental Clinical Specialties. Madrid. Spain
Med. oral patol. oral cir. bucal (Internet) ; 23(5): e602-e607, sept. 2018. ilus, tab, graf
Artículo en Inglés | IBECS | ID: ibc-176380
Biblioteca responsable: ES1.1
Ubicación: BNCS
ABSTRACT

BACKGROUND:

A correct design is needed in short implants to improve primary stability (PS) in low quality bone. This study aimed to compare PS of double thread and single thread short implants. MATERIAL AND

METHODS:

Thirty implants with single thread design (PHI/SHORT-I) and 30 implants with double thread design (PHIA/SHORT-I) (Radhex(R), Inmet-Garnick S.A., Guadalajara, Spain) were placed in 30 randomly selected bovine ribs. PS was assessed in implant stability quotients (ISQ) and periotest values (PV) with Osstell(TM) and Periotest(R) devices, respectively. Computed tomographies of the ribs were taken and bone quality was evaluated in Hounsfield Units (HU) using Ez3D Plus software (Vatech Co., Korea). Only implants placed in low quality bone according to Misch and Kircos classification were selected (D3 bone 350-850 HU; and D4 bone 150-350 HU). Ten implants were not included in the study for being placed in D1 and D2 bone. Finally, 50 implants were selected 17 and 9 PHI/SHORT-I in D3 and D4 bone respectively, and 15 and 9 PHIA/SHORT-I in D3 and D4 bone respectively.

RESULTS:

The one-way ANOVA showed statistically significant differences in ISQ (61.35 ± 4.77 in PHI/SHORT-I and 66.43 ± 4.49 in PHIA/SHORT-I, P < 0.005) and PV (-2.76 ± 0.8 and -4.11 ± 1.24 respectively, P < 0.005) between two implant designs in D3 bone, and statistically significant differences in ISQ (53.44 ± 3.34 in PHI/SHORT-I and 60.56 ± 1.53 in PHIA/SHORT-I, P<0.0001) and PV (1.13 ± 0.95 and -2.5 ± 0.61 respectively, P < 0.0001) between two groups in D4 bone.

CONCLUSIONS:

Double thread design short implants resulted to have higher PS in comparison with single thread design short implants in D3 and D4 bone
Asunto(s)

Texto completo: Disponible Colección: Bases de datos nacionales / España Base de datos: IBECS Asunto principal: Retención de Dentadura / Implantación Dental Límite: Humanos Idioma: Inglés Revista: Med. oral patol. oral cir. bucal (Internet) Año: 2018 Tipo del documento: Artículo Institución/País de afiliación: Complutense University/Spain

Texto completo: Disponible Colección: Bases de datos nacionales / España Base de datos: IBECS Asunto principal: Retención de Dentadura / Implantación Dental Límite: Humanos Idioma: Inglés Revista: Med. oral patol. oral cir. bucal (Internet) Año: 2018 Tipo del documento: Artículo Institución/País de afiliación: Complutense University/Spain
...