Valoración de sueros con bajo índice S/CO utilizando 2 sistemas de quimioluminiscencia para la detección de anticuerpos frente al virus de la hepatitis C y su correlación con la detección de ARN viral / Evaluation of sera with a low signal to cut-off ratio using two chemiluminescent assays for detecting Hepatitis C Virus, and their correlation with the detection of Viral RNA
López-Fabal, M Fátima; Pérez-Rivilla, Alfredo; Gómez-Garcés, José Luis.
Afiliación
  • López-Fabal, M Fátima; Hospital Universitario de Móstoles. Servicio de Microbiología. Móstoles. España
  • Pérez-Rivilla, Alfredo; Hospital 12 de Octubre. Servicio de Microbiología. Madrid. España
  • Gómez-Garcés, José Luis; Hospital Universitario de Móstoles. Servicio de Microbiología. Móstoles. España
Enferm. infecc. microbiol. clín. (Ed. impr.) ; 36(4): 222-224, abr. 2018. tab
Article en Es | IBECS | ID: ibc-176515
Biblioteca responsable: ES1.1
Ubicación: BNCS
RESUMEN
OBJETIVO: Los ensayos comerciales utilizados para demostrar la presencia de anticuerpos anti-VHC establecen, cada uno de ellos, puntos de corte determinados que categorizan los resultados en función de los mismos, aunque el problema de los resultados falsos positivos en el cribado de sueros de hepatitis C sea bien conocido. El objetivo de este trabajo ha sido valorar los resultados obtenidos por 2 ensayos quimioluminiscentes en una serie de sueros seleccionados, contrastando estos resultados con la detección de ARN viral en las muestras estudiadas. MATERIAL Y MÉTODOS: Se seleccionaron 200 sueros considerados como reactivos (positivos), aunque con un bajo índice S/CO, utilizando 2 ensayos de quimioluminiscencia y posteriormente fueron sometidos a amplificación genómica. RESULTADOS Y DISCUSIÓN: Solamente en 8 (4%) de las muestras seleccionadas pudo detectarse ARN viral. A la vista de estos resultados, consideramos que el diseño de los ensayos de quimioluminiscencia empleados no ofrecen una especificidad suficiente utilizados como pruebas únicas para el diagnóstico de la hepatitis C
ABSTRACT
OBJECTIVE: All commercial assays used to measure the presence of Hepatitis C virus (HCV) antibodies set cut-off points to categorise the results, but the problem of false positive results in screening hepatitis C sera is well known. The aim of this study was to evaluate the results obtained by two chemiluminescent assays in selected sera, and compare these results with the detection of viral RNA in the specimens studied. MATERIAL AND METHODS: Two hundred reactive sera (positive) were selected, although with a low signal to cut-off ratio (S/CO), were selected, using two chemiluminescent assays and were then subjected to genome amplification. RESULTS AND DISCUSSION: Viral RNA could be only be detected in 8 (4%) of the selected specimens. Taking these results into account, we believe that the design of the current chemiluminescent assays do not provide sufficient specificity when they are used as the only tests for the diagnosis of hepatitis C