Your browser doesn't support javascript.
loading
Comparación de la corrección de la discrepancia de longitud de miembros tras el empleo de un vástago con cuello modular y su homólogo monobloque en la artroplastia total de cadera primaria / Comparison of leg length discrepancy correction after the use of a modular neck stem and its monoblock homologue in total primary hip arthroplasty
López, R. E; Tomás, J. M Pelayo de; Morales Suárez Varela, M; Rodrigo Pérez, J. L.
Afiliación
  • López, R. E; Hospital Universitario Doctor Peset. España
  • Tomás, J. M Pelayo de; Hospital Universitario Doctor Peset. Universidad de Valencia. Valencia. España
  • Morales Suárez Varela, M; Universidad de Valencia. Facultad de Medicina. Facultad de Farmacia. Valencia. España
  • Rodrigo Pérez, J. L; Hospital Universitario Doctor Peset. Universidad de Valencia. Valencia. España
Rev. esp. cir. ortop. traumatol. (Ed. impr.) ; 66(6): 445-453, Nov-Dic. 2022. ilus, tab
Artículo en Español | IBECS | ID: ibc-210655
Biblioteca responsable: ES1.1
Ubicación: ES15.1 - BNCS
RESUMEN
Introducción y

objetivos:

Los vástagos con doble modularidad buscan restaurar de forma más precisa la anatomía al permitir ajustes intraoperatorios gracias a los cuellos modulares. Nuestro objetivo es comparar la corrección radiográfica de la longitud con el vástago H MAX-M® frente a su homólogo monobloque H MAX-S®. Material y

métodos:

Se realizó un estudio de cohortes prospectivo mediante muestreo consecutivo sobre pacientes intervenidos de artroplastia total de cadera primaria con el diagnóstico de coxartrosis entre el año 2011 y 2015. Un brazo de la cohorte incluyó a los pacientes intervenidos con vástago modular y el otro con vástago monobloque. Se midió la longitud en la radiografía anteroposterior de pelvis en carga a los seis meses. Las medias de las mediciones obtenidas para cada brazo de la cohorte se compararon entre sí.

Resultados:

No se han observado diferencias estadísticamente significativas en la corrección de la disimetría entre ambos grupos determinada como la diferencia de longitud entre la cadera operada y la cadera contralateral (p=0,106). Tampoco se observaron diferencias en los valores postoperatorios de longitud (p=0,053). Cabe decir que tanto para el vástago modular como para el vástago monobloque el grupo mayoritario es aquel con longitud restaurada (84,1 y 80,4%, respectivamente; p=0,001).

Conclusión:

A pesar de la ventaja teórica de la modularidad y que disponer de piezas intercambiables podría ser de gran interés, en nuestro estudio no hemos podido demostrar que exista una superioridad de los diseños modulares frente al monoblock para control de la disimetría postoperatoria.(AU)
ABSTRACT
Introduction and

objective:

Dual modularity stems seek to more precisely restore anatomy by allowing intraoperative adjustments thanks to modular necks. Our aim is to compare the radiographic length correction with the H MAX-M® Stem versus its monoblock counterpart H MAX-S®. Material and

methods:

A prospective cohort study was carried out through consecutive sampling on patients who underwent primary total hip arthroplasty with coxarthrosis diagnosis between 2011 and 2015. One arm of the cohort included patients who were operated with a modular stem and the other with a monobloc stem. Length was measured on the anteroposterior pelvic-bearing radiograph at six months. The mean of the measurements obtained for each arm of the cohort were compared with each other.

Results:

No statistically significant differences were observed in the correction of asymmetry between both groups, determined as the difference in length between the operated hip and the contralateral hip (P=.106). Nor were differences observed in postoperative length values (P=.053). It should be noted that for both the modular stem and the monobloc stem, the majority group is the one with restored length (84.1% and 80.4%, respectively; P=.001).

Conclusion:

Despite the theoretical advantage of modularity and that having interchangeable parts could be of great interest, in our study, we have not been able to demonstrate a superiority of modular designs compared to monoblock for control of postoperative leg length discrepancy.(AU)
Asunto(s)


Texto completo: Disponible Colección: Bases de datos nacionales / España Base de datos: IBECS Asunto principal: Artroplastia / Osteoartritis de la Cadera / Ataxia Cerebelosa / Lesiones de la Cadera / Cadera Límite: Femenino / Humanos / Masculino Idioma: Español Revista: Rev. esp. cir. ortop. traumatol. (Ed. impr.) Año: 2022 Tipo del documento: Artículo Institución/País de afiliación: Hospital Universitario Doctor Peset/España / Universidad de Valencia/España

Texto completo: Disponible Colección: Bases de datos nacionales / España Base de datos: IBECS Asunto principal: Artroplastia / Osteoartritis de la Cadera / Ataxia Cerebelosa / Lesiones de la Cadera / Cadera Límite: Femenino / Humanos / Masculino Idioma: Español Revista: Rev. esp. cir. ortop. traumatol. (Ed. impr.) Año: 2022 Tipo del documento: Artículo Institución/País de afiliación: Hospital Universitario Doctor Peset/España / Universidad de Valencia/España
...