Comparison of the fracture torque of different Brazilian mini-implants
Lima, Guilherme Machado Alvares de; Soares, Mário Sérgio; Penha, Sibele Sarti; Romano, Marcelo Munhóes.
Afiliación
  • Lima, Guilherme Machado Alvares de; University of São Paulo. School of Dentistry. São Paulo. BR
  • Soares, Mário Sérgio; University of São Paulo. School of Dentistry. Department of Stomatology. Integrated Clinic. São Paulo. BR
  • Penha, Sibele Sarti; University of São Paulo. School of Dentistry. Department of Stomatology. Integrated Clinic. São Paulo. BR
  • Romano, Marcelo Munhóes; University of São Paulo. School of Dentistry. Department of Stomatology. Integrated Clinic. São Paulo. BR
Braz. oral res ; 25(2): 116-121, Mar.-Apr. 2011. ilus, tab
Article en En | LILACS | ID: lil-583848
Biblioteca responsable: BR1.1
ABSTRACT
This study evaluated fracture torque by torsion, in relation to the length and diameter of orthodontic mini-implants, to demonstrate their viability for clinical and experimental use based on the torque recommended by the manufacturers. The fractures at the moment of insertion, whose incidence in the literature is around 4 percent, are principally due to excessive force and the inability of the implant to resist rotational forces. Thirty orthodontic mini-implants of three commercial brands available in Brazil (Neodent 1.6 x 9 mm, Dentoflex 1.6 x 9 mm and Kopp 1.6 x 9 mm) were attached to a device made specifically for this research, leaving the mini-implants with sufficient stability. The miniimplants were submitted to torsion torque, using a digital torque wrench, until their breaking point. The values obtained with the test were submitted to analysis of variance and the Tukey test. The mean values of mini-implant ruptures were 26 N.cm for group A (Dentoflex), 25.4 N. cm for group B (Kopp) and 32.8 N.cm for group C (Neodent). From the Tukey test we could observe that the relationships between the means of the Dentoflex and Neodent groups, and between the Kopp and Neodent groups, were significant. Between the Dentoflex and Kopp groups, significance was nonexistent. All the values found in our research for fracture torque were higher than the limits recommended by the manufacturers for clinical use in orthodontics. The highest values were found in the Neodent group.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: LILACS Asunto principal: Titanio / Implantes Dentales / Torque / Métodos de Anclaje en Ortodoncia País/Región como asunto: America do sul / Brasil Idioma: En Revista: Braz. oral res Asunto de la revista: ODONTOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Brasil

Texto completo: 1 Colección: 01-internacional Base de datos: LILACS Asunto principal: Titanio / Implantes Dentales / Torque / Métodos de Anclaje en Ortodoncia País/Región como asunto: America do sul / Brasil Idioma: En Revista: Braz. oral res Asunto de la revista: ODONTOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Brasil