Implante por cateter de bioprótese valvar para tratamento da estenose aórtica: experiência de três anos / Transcatheter bioprosthesis implantation for the treatment of aortic stenosis: three-year experience
Brito Junior, Fabio Sandoli de; Abizaid, Alexandre; Almeida, Breno O.; Caixeta, Adriano; Tarasoutchi, Flávio; Grube, Eberhard; Perin, Marco.
Afiliación
  • Brito Junior, Fabio Sandoli de; Hospital Israelita Albert Einstein. São Paulo. BR
  • Abizaid, Alexandre; Hospital Israelita Albert Einstein. São Paulo. BR
  • Almeida, Breno O.; Hospital Israelita Albert Einstein. São Paulo. BR
  • Caixeta, Adriano; Hospital Israelita Albert Einstein. São Paulo. BR
  • Tarasoutchi, Flávio; Hospital Israelita Albert Einstein. São Paulo. BR
  • Grube, Eberhard; Hospital Israelita Albert Einstein. São Paulo. BR
  • Perin, Marco; Hospital Israelita Albert Einstein. São Paulo. BR
Arq. bras. cardiol ; 99(2): 697-705, ago. 2012. graf, tab
Artículo en Portugués | LILACS | ID: lil-647716
Biblioteca responsable: BR1.1
RESUMO
FUNDAMENTO O implante por cateter de bioprótese valvar aórtica é uma nova modalidade de tratamento para portadores de estenose aórtica inoperáveis ou de alto risco cirúrgico.

Objetivo:

Relatar a experiência de três anos do implante por cateter da bioprótese CoreValve.

MÉTODOS:

Entre janeiro de 2008 e janeiro de 2011, 35 pacientes com estenose aórtica (33 casos) ou disfunção de bioprótese valvar aórtica (dois casos) de alto risco cirúrgico foram submetidos ao implante da bioprótese CoreValve.

RESULTADOS:

A média de idade dos pacientes foi 81,5 ± 9 anos, e 80% apresentavam-se em classe funcional III ou IV de insuficiência cardíaca. O EuroScore foi 18,4 ± 14,3% e o STS 14,5 ± 11,6%. Obteve-se sucesso do implante em 34 (97,1%) pacientes. Após a intervenção houve redução do gradiente transvalvar de 84,9 ± 22 para 22,5 ± 9,5 mmHg e 87,1% dos pacientes evoluíram em classe funcional I ou II. A mortalidade aos 30 dias e no seguimento médio de 400 ± 298 dias foi, respectivamente, de 11,4% e 31,4%. A ocorrência de complicações hemorrágicas com risco de morte foi o único preditor independente de mortalidade cardiovascular. Acidente vascular cerebral ocorreu em 5,7% dos pacientes. Marca-passo permanente foi necessário em 32,1% dos casos no primeiro mês após o procedimento.

CONCLUSÃO:

O implante por cateter de bioprótese valvar aórtica é um procedimento seguro e eficaz para ser empregado em portadores de estenose aórtica de alto risco cirúrgico. O dispositivo CoreValve é eficaz no médio-prazo, em seguimento de até três anos.
ABSTRACT

BACKGROUND:

Transcatheter aortic bioprosthesis implantation is a new treatment modality for patients with aortic stenosis who are inoperable or at high surgical risk.

OBJECTIVE:

To report the three-year experience with transcatheter CoreValve® bioprosthesis implantation.

METHODS:

From January 2008 to January 2011, 35 patients with aortic stenosis (33) or aortic valve bioprosthesis dysfunction (two) at high surgical risk underwent transcatheter CoreValve® bioprosthesis implantation.

RESULTS:

The patients' mean age was 81.5 ± 9 years, and 80% had heart failure functional class III or IV. The EuroScore was 18.4 ± 14.3% and the STS risk score was 14.5 ± 11.6%. Successful device implantation was achieved in 34 (97.1%) patients. After the intervention, a reduction in the transvalvular pressure gradient from 84.9 ± 22 to 22.5 ± 9.5 mm Hg was observed, and 87.1% of the patients progressed to functional class I or II. Thirty-day mortality and mortality in the follow-up of 400 ± 298 days were 11.4% and 31.4%, respectively. The occurrence of life-threatening hemorrhagic complications was the only independent predictor of cardiovascular mortality. Stroke occurred in 5.7% of the patients. Permanent pacemaker was required in 32.1% of the patients within the first month after the procedure.

CONCLUSION:

Transcatheter aortic bioprosthesis implantation is a safe and effective procedure to be used in patients with aortic stenosis at high surgical risk. The CoreValve® prosthesis proved to have mid-term efficacy in a three-year follow-up.
Asunto(s)


Texto completo: Disponible Colección: Bases de datos internacionales Contexto en salud: ODS3 - Meta 3.4 Reducir las muertes prematuras por enfermedades no transmisibles Problema de salud: Enfermedad Cardiovascular Base de datos: LILACS Asunto principal: Estenosis de la Válvula Aórtica / Bioprótesis / Prótesis Valvulares Cardíacas / Cateterismo Cardíaco / Implantación de Prótesis de Válvulas Cardíacas Tipo de estudio: Estudio pronóstico Límite: Anciano / Anciano de 80 o más años / Femenino / Humanos / Masculino Idioma: Portugués Revista: Arq. bras. cardiol Asunto de la revista: Cardiología Año: 2012 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Hospital Israelita Albert Einstein/BR

Texto completo: Disponible Colección: Bases de datos internacionales Contexto en salud: ODS3 - Meta 3.4 Reducir las muertes prematuras por enfermedades no transmisibles Problema de salud: Enfermedad Cardiovascular Base de datos: LILACS Asunto principal: Estenosis de la Válvula Aórtica / Bioprótesis / Prótesis Valvulares Cardíacas / Cateterismo Cardíaco / Implantación de Prótesis de Válvulas Cardíacas Tipo de estudio: Estudio pronóstico Límite: Anciano / Anciano de 80 o más años / Femenino / Humanos / Masculino Idioma: Portugués Revista: Arq. bras. cardiol Asunto de la revista: Cardiología Año: 2012 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Hospital Israelita Albert Einstein/BR
...