Your browser doesn't support javascript.
loading
Diseño de un panel multicolor para evaluar moléculas intracelulares y de superficie mediante citometría de flujo / Design of a multicolor panel to assess intracellular and surface molecules by flow cytometry
Mateus, Jose; Lasso, Paola; González, John Mario; Puerta, Concepción Judith; Cuéllar, Adriana.
Afiliación
  • Mateus, Jose; Pontificia Universidad Javeriana. Grupo de Inmunobiología y Biología Celular. Bogotá. CO
  • Lasso, Paola; Pontificia Universidad Javeriana. Grupo de Inmunobiología y Biología Celular. Bogotá. CO
  • González, John Mario; Pontificia Universidad Javeriana. Grupo de Inmunobiología y Biología Celular. Bogotá. CO
  • Puerta, Concepción Judith; Pontificia Universidad Javeriana. Grupo de Inmunobiología y Biología Celular. Bogotá. CO
  • Cuéllar, Adriana; Pontificia Universidad Javeriana. Grupo de Inmunobiología y Biología Celular. Bogotá. CO
Biomédica (Bogotá) ; Biomédica (Bogotá);33(4): 660-672, Dec. 2013. ilus, graf, tab
Article en Es | LILACS | ID: lil-700482
Biblioteca responsable: CO332
RESUMEN
Introducción. La citometría de flujo permite detectar la presencia de moléculas intracelulares y de superficie, de forma simultánea sobre cada célula. Objetivo. Describir un método para la construcción armónica de un panel multicolor con 11 parámetros para el análisis fenotípico y funcional de linfocitos T (LT) CD8 + por citometría de flujo. Materiales y métodos. Para la construcción del panel multicolor, se seleccionaron las moléculas y se titularon los conjugados con fluorocromos para la determinación de CD3, CD8, CCR7, CD28, CD27, CD45RA, CD95 y CD127, en células mononucleares de sangre periférica. Para la evaluación del panel, se hizo la construcción progresiva adicionando uno a uno los conjugados y la fluorescencia menos uno (FMO). Este método fue aplicado para células ex vivo y para evaluar la producción de IFN ? , IL-2 y TNFa frente al estímulo con la enterotoxina B de Staphylococcus aureus (SEB) y al antígeno crudo de Trypanosoma cruzi . Finalmente, se procedió al análisis de las subpoblaciones de LT CD8 + ex vivo en individuos sanos. Resultados. La evaluación de las moléculas con los conjugados no mostró interferencia en las señales de fluorescencia. Las frecuencias de las subpoblaciones de LT CD8 + evaluadas fueron cercanas a los valores reportados en otros estudios. Además, se observó que la frecuencia de LT CD8 + productores de IFN ? , IL-2 y TNFa fue mayor a las seis horas de cultivo con SEB y con el antígeno crudo de T. cruzi . Conclusiones. El método aplicado para la construcción del panel multicolor permite obtener frecuencias de las subpoblaciones de LT CD8 + que corresponden a lo reportado en la literatura científica.
ABSTRACT

Introduction:

Flow cytometry allows simultaneous detection of surface and intracellular molecules on each cell.

Objective:

To describe a method for building up a harmonic multicolor panel with 11 flow cytometry parameters for phenotypic and functional analysis on CD8 + T lymphocytes. Materials and

methods:

For the multicolor panel construction, we selected the molecules and titred conjugated antibodies with fluorochromes for CD3, CD8, CCR7, CD28, CD27, CD45RA, CD95 and CD127 determination in peripheral blood mononuclear cells (PBMC). To evaluate the panel, the conjugated antibodies were gradually added one by one and fluorescence minus one (FMO) test was performed. This method was applied to assess ex vivo subpopulations of T cells and the production of intracellular IFN ? , IL-2 and TNF a using polyclonal stimulation with enterotoxin B from Staphylococcus aureus (SEB) and antigen-specific cells with crude Trypanosoma cruzi antigen. Finally, the ex vivo CD8 + T lymphocyte subpopulations frequencies were analyzed in healthy individuals.

Results:

The evaluation of the selected molecules and conjugates did not show interference in the fluorescence signals and detection. The frequencies of CD8 + T cells evaluated were similar to the values reported in other studies. Additionally, we observed that the frequency of CD8 + T lymphocytes producing IFN ? , IL-2 and TNF a was higher 6 hours after culture with SEB and crude T. cruzi lysate.

Conclusions:

The method used for the construction of a multicolor panel allows obtaining frequencies of CD8 + T lymphocyte subpopulations corresponding to those reported in the literature.
Asunto(s)
Palabras clave
Texto completo: 1 Colección: 01-internacional Base de datos: LILACS Asunto principal: Citocinas / Subgrupos de Linfocitos T / Citometría de Flujo Límite: Humans Idioma: Es Revista: Biomédica (Bogotá) Asunto de la revista: MEDICINA Año: 2013 Tipo del documento: Article País de afiliación: Colombia Pais de publicación: Colombia
Texto completo: 1 Colección: 01-internacional Base de datos: LILACS Asunto principal: Citocinas / Subgrupos de Linfocitos T / Citometría de Flujo Límite: Humans Idioma: Es Revista: Biomédica (Bogotá) Asunto de la revista: MEDICINA Año: 2013 Tipo del documento: Article País de afiliación: Colombia Pais de publicación: Colombia