T2* and proton density measurement of normal human lung parenchyma using submillisecond echo time gradient echo magnetic resonance imaging.
Eur J Radiol
; 29(3): 245-52, 1999 Mar.
Article
en En
| MEDLINE
| ID: mdl-10399610
ABSTRACT
OBJECTIVE:
To obtain T2* and proton density measurements of normal human lung parenchyma in vivo using submillisecond echo time (TE) gradient echo (GRE) magnetic resonance (MR) imaging. MATERIALS ANDMETHODS:
Six normal volunteers were scanned using a 1.5-T system equipped with a prototype enhanced gradient (GE Signa, Waukausha, WI). Images were obtained during breath-holding with acquisition times of 7-16 s. Multiple TEs ranging from 0.7 to 2.5 ms were tested. Linear regression was performed on the logarithmic plots of signal intensity versus TE, yielding measurements of T2* and proton density relative to chest wall muscle. Measurements in supine and prone position were compared, and effects of the level of lung inflation on lung signal were also evaluated.RESULTS:
The signal from the lung parenchyma diminished exponentially with prolongation of TE. The measured T2* in six normal volunteers ranged from 0.89 to 2.18 ms (1.43 +/- 0.41 ms, mean +/- S.D.). The measured relative proton density values ranged between 0.21 and 0.45 (0.29 +/- 0.08, mean +/- S.D.). Calculated T2* values of 1.46 +/- 0.50, 1.01 +/- 0.29 and 1.52 +/- 0.18 ms, and calculated relative proton densities of 0.20 +/- 0.03, 0.32 +/- 0.13 and 0.35 +/- 0.10 were obtained from the anterior, middle and posterior portions of the supine right lung, respectively. The anterior-posterior proton density gradient was reversed in the prone position. There was a pronounced increase in signal from lung parenchyma at maximum expiration compared with maximum inspiration. The ultrashort TE GRE technique yielded images demonstrating signal from lung parenchyma with minimal motion-induced noise.CONCLUSION:
Quantitative in vivo measurements of lung T2* and relative proton density in conjunction with high-signal parenchymal images can be obtained using a set of very rapid breath-hold images with a recently developed ultrashort TE GRE sequence.