Your browser doesn't support javascript.
loading
Enzyme amplification as detection tool in continuous-flow systems. I. Development of an enzyme-amplified biochemical detection system coupled on-line to flow-injection analysis.
van Bommel, M R; de Jong, A P; Tjaden, U R; Irth, H; van der Greef, J.
Afiliación
  • van Bommel MR; Division of Analytical Chemistry, Leiden/Amsterdam Center of Drug Research, Leiden University, The Netherlands.
J Chromatogr A ; 855(2): 383-96, 1999 Sep 10.
Article en En | MEDLINE | ID: mdl-10519082
ABSTRACT
The on-line coupling of flow-injection analysis (FIA) to an enzyme-amplified biochemical detection (EA-BCD) system is described. The aim of this study is the development of a detection system able to detect biotin-containing compounds at low concentration levels. The detection system is based on the interaction of biotin with enzyme-labeled affinity proteins. Biotin possesses a high affinity to both streptavidin and anti-biotin Fab fragments, which are both tested. Several biotin derivatives are available with different reactive probes, which can be used to label analytes of interest. Therefore, biotin acts as a universal probe for the enzyme-amplified biochemical detection. Alkaline phosphatase (AP) was used as enzyme label. Several parameters, such as substrate type and concentration, concentration of enzyme-labeled affinity protein, reaction time and reaction temperature were examined. Biotin aminocaproic acid was used as a model compound. In addition to biotin aminocaproic hydrazide, other biotinylation reagents were also examined. With fluorescence detection of the enzyme-generated product, a mass detection limit of 1 fmol was achieved.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Análisis de Inyección de Flujo / Enzimas Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Revista: J Chromatogr A Año: 1999 Tipo del documento: Article País de afiliación: Países Bajos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Análisis de Inyección de Flujo / Enzimas Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Revista: J Chromatogr A Año: 1999 Tipo del documento: Article País de afiliación: Países Bajos