[Slice sensitivity profile and image pixel noise of multi-slice spiral CT in comparison with single slice spiral CT]. / Schichtempfindlichkeitsprofile und Bildpunktrauschen einer Mehrschicht Spiral-CT im Vergleich zu einer Einzelschicht Spiral-CT.
Rofo
; 171(3): 219-25, 1999 Sep.
Article
en De
| MEDLINE
| ID: mdl-10520332
PURPOSE: Presentation and evaluation of slice sensitivity profile and pixel noise of multi-slice CT in comparison to single-slice CT. METHODS: Slice sensitivity profiles and pixel noise of a multi-slice CT equipped with a 2D matrix detector array and of a single-slice CT were evaluated in phantom studies. RESULTS: For the single-slice CT the width of the slice sensitivity profiles increased with increasing pitch. In spite of a much higher table speed the slice sensitivity profiles of multi-slice CT were narrower and did not increase with higher pitch. Noise in single-slice CT was independent of pitch. For multi-slice CT noise increased with higher pitch and for the higher pitch decreased slightly with higher detector row collimation. CONCLUSIONS: Multi-slice CT provides superior z-resolution and higher volume coverage speed. These qualities fulfill one of the prerequisites for improvement of 3D postprocessing.